dx.doi.org/10.14227/DT140407P46

Question and Answer Section - November 2007

William Brown and Margareth Marques
The following questions have been submitted by readers of Dissolution Technologies. Margareth Marques, Ph. D. and Will Brown, United States Phamacopeia, authored responses to each of the questions.
*Note: These are opinions and interpretations of the authors, and are not necessarily the official viewpoints of the USP

Email for correspondence: web@usp.org

Q At the time of sampling in a dissolution test, is it appropriate to increase the speed of the paddle to 250 rpm for a few seconds in order to get better homogenization?
A No, the speed of the basket or paddle must remain constant during the entire dissolution test. The USP General Chapter <711> Dissolution, on page 278 of USP 30, states, “A speed-regulating device is used that allows the shaft rotation speed to be selected and maintained at the rate specified in the individual monograph, within ± 4%.” In addition, the discriminative power of the test will be lost at such a high speed. The expectation of the dissolution test is to determine the release rate under controlled conditions. It is not an assay of the contents where total recovery of the drug from the dosage unit is the goal.

Q Are there any justifications for getting lower results at the later time points when doing a dissolution profile?
A Possible reasons are (1) the calculation formula may not be appropriate, a correction must be done to account for the volume of medium removed and replaced and the amount of analyte removed; (2) the drug substance may not be stable in the dissolution medium and is degrading during the test; and (3) the concentration of the drug substance released to the medium may be close to the saturation concentration and the drug substance still present in the dosage form is not being solubilized.

Q Where can I find information about running dissolution or drug release tests for suppositories?
A The use of suppositories is more popular in Europe than in the United States. The European Pharmacopoeia has a general chapter, 2.9.42 Dissolution for Lipophilic Solid Dosage Forms, that describes an apparatus, a flowthough cell (USP Apparatus 4) with a special cell, and the method that can be used to evaluate the drug release from suppositories. Lipophilic and hydrophilic suppositories will typically require different test conditions. Also, you can find papers in the literature with more information at the site www.NCBI.NLM.NIH.GOV (access free of charge). This is a chapter on suppositories in the new dissolution book out called “Dissolution Theory, Methodology and Testing,“ Edited by A. Palmieri�see ad on page 32.

Q Some dissolution equipment are fitted with eight dissolution vessels. Is it possible to do dissolution runs in this type of equipment combining different product batches? As an example, could six vessels be used with one batch and the other two vessels with the second batch?
A The procedure described for dissolution does not specifically forbid testing more (or less) than six units at a time. In USP, a procedure for one unit is given followed by the statement, “Repeat the test with additional dosage form units.” A consideration of the effect of intermediate precision on the results should be given for units within the same sample (e.g., the six units tested) when the sample is divided between two actual runs as in the question, where one lot is represented with all six tested at the same time while the second lot would only have two units in that run with the remaining four tested in the following run. Whatever course is decided, standard procedures should be in place as a guide.

Q Can nitrogen gas be introduced in the dissolution vessel during the test in order to degas the medium?
A If nitrogen is introduced to the medium during testing, it will need to be brought in through a tube. The presence of a tube in the dissolution test during testing will change the hydrodynamics of the test. Additionally, if the nitrogen exits the medium as a stream of bubbles, additional unwanted agitation will result.

As a separate issue, if the dissolved gas content of the medium is controlled using nitrogen, the use of a dissolved oxygen meter to confirm the dissolved gas content is not recommended. The nitrogen will itself saturate the medium while preferentially removing dissolved oxygen.

Q During the validation of dissolution methods, should the items specificity, accuracy, and precision be carried out in dissolution vessels or in volumetric flasks?
A It is advisable to run the specificity, accuracy, and precision evaluations according to the dissolution procedure, which means in the dissolution vessel. You need to replicate the conditions of agitation and temperature to which the product will be submitted. Regarding precision, you will evaluate repeatability and intermediate precision, and in both cases, it means running the test in the conditions specified in the dissolution method to evaluate the variability associated with the test and with the laboratory procedures and conditions.

Q How many dosage forms units should be used to evaluate the intermediate precision of a dissolution procedure?
A At least six units should be used. Typically, laboratories are doing the evaluation of intermediate precision in duplicate, with 12 units.

Q In the USP General Chapter <1225> Validation of Compendial Procedures and in the ICH guidance Q2(R1), it is recommended to evaluate only precision in the case of dissolution procedures. Is this enough?
A Both documents have a footnote stating that other items may be required, depending on the nature of the test. Because there were no official guidances on how to select which parameters should be validated and how to do it, USP developed a new general chapter, <1092> The Dissolution Procedure: Development and Validation. This general chapter gives some recommendations on which characteristics should be evaluated in the validation of dissolution procedures.