dx.doi.org/10.14227/DT150208P37

Question and Answer Section - May 2008

William Brown and Margareth Marques
The following questions have been submitted by readers of Dissolution Technologies. Margareth Marques, Ph. D. and Will Brown, United States Phamacopeia, authored responses to each of the questions.
*Note: These are opinions and interpretations of the authors, and are not necessarily the official viewpoints of the USP

Email for correspondence: web@usp.org

Q Could you provide some guidelines on how to do validation of dissolution methods?
A The ICH guidance Q2(R1), Validation of Analytical Procedures: Text and Methodology (available at www.ich.org), does not give details on how to validate dissolution tests. The U.S. Pharmacopeia developed a new general chapter, <1092> The Dissolution Procedure: Development and Validation, where more details on what parameters should be validated and how to do this validation are provided. The new general chapter was preceded by a paper that contains suggestions on acceptance criteria for the validation parameters (see Gray, V. A.; Brown, C. K.; Dressman, J. B.; Leeson, L. J. A new general information chapter on dissolution. Pharm. Forum 2001, 27 (6), 3432�3439).

Q Can the linearity in the validation of dissolution methods be done using volumetric glassware instead of dissolution vessels?
A Yes. You are going to prepare at least five standard solutions of the drug substance, ranging in concentration from approximately ±20% of the lowest to ±20% of the highest concentration expected to be released in the dissolution profile. If the solubility of the drug substance in the dissolution medium is not complete, organic solvents at no more than 5% (v/v) of the final volume of the first dilution can be used to enhance the solubility of the compound.

Q How is the accuracy of a dissolution method evaluated?
A To evaluate the accuracy of dissolution procedures, it is recommended that all the components of the formulation, in the same proportion as found in the dosage form, are transferred to the dissolution vessel. This step can be done by weighing each component of the formulation separately or by using a placebo blend. A minimum of three concentration levels of the drug, ranging from approximately ±20% of the lowest to ±20% of the highest concentration expected to be released in the dissolution profile, should be evaluated. In the case of low solubility of the drug substance in the dissolution medium, an organic solvent not exceeding 5% (v/v) of the first dilution can be used. An aliquot of a stock solution of the drug substance in organic solvent can be transferred to the vessel, or the drug substance can be transferred to the vessel and dissolved with organic solvent, then all the other components of the formulation are added. In any case, the amount of organic solvent used should not exceed 5% of the volume in the vessel.

Q When the USP deaeration procedure is used, the medium ends up at 41 °C. How can the volume specified for the test be measured for transfer to the dissolution vessel?
A Considering the difference in density of the medium at 41 °C from that at 25 °C, a larger volume of the heated, deaerated medium needs to be transferred than is specified for the test. To compensate for this difference, the USP laboratory determines the density at 25 °C, and then by calculating the mass equivalent to the specified volume, the appropriate weight can be measured out into each vessel. Please note that the USP General Chapter states that the volume given in the test procedure refers to measurements made between 20 and 25 °C.

Q Can the replacement of the medium in the dissolution vessel create turbulence?
A Yes, when replacing the medium withdrawn during a dissolution profile or dissolution test for delayed- or extended-release dosages, turbulence is going to be an issue. Therefore, this procedure needs to be carried out as gently as possible. One possible way is to slowly deliver the volume of medium onto the wall of the vessel using a pipet or a syringe.

Q Does the use of a hollow shaft for the sampling meet the USP sampling criteria?
A Every time a sampling procedure that deviates from the procedure described in the USP General Chapter <711> Dissolution is going to be used, it is necessary to demonstrate that the alternative procedure is equivalent to the USP procedure, and it must be validated.

Q Can baskets with other mesh measurements be used in dissolution testing?
A The most used basket is 40-mesh, but baskets with other mesh sizes can be used if properly justified. The alternative mesh should be properly characterized in the documentation for the procedure.

Q What should be the specific distance between two dissolution apparatus on the same bench?
A A measurement of the vibration at the dissolution equipment should be made with all the equipment on the same bench running. Then, the positioning of the equipments on the bench can be modified until the lowest vibration level is found. The whole lab environment needs to be taken into account because vibration can come from other areas close to the lab through the floor or walls. The vibration should be measured in several conditions, and an attempt to minimize it should be made. A discussion of the possible acceptable levels of vibration at the dissolution test assembly can be found in the Dissolution Toolkit available at the USP website, http://www.usp.org/USPNF/compendialTools.html

Q Should the pH of the dissolution medium be adjusted before heating or after heating?
A The pH of the dissolution medium should be adjusted at room temperature, before heating it in the dissolution bath.

Q Most of the dissolution methods employ 500�900 mL of dissolution medium, but the patient usually takes a tablet or a capsule with about 250 mL of water or juice. How can these differences be justified?
A The volume of medium used in the dissolution test is defined by taking into account the sink conditions of the drug substance being analyzed. The solubility of the drug substance in various media is determined, the saturation concentration is calculated considering the highest dose of the product, and the volume of medium is defined by at least three times the volume of a saturated solution. The dissolution test is not necessarily limited to in vivo conditions. The volume of 250 mL of medium is considered when classifying the solubility of drug substances according to the Biopharmaceutics Classification System (see FDA Guidance for Industry�Waiver of In Vivo Bioavailability and Bioequivalence Studies for Immediate- Release Solid Oral Dosage Forms Based on a Biopharmaceutics Classification System, available at www.fda.gov/cder/guidance/3618fnl.htm). According to this guidance, a drug substance is considered highly soluble when the highest dose strength is soluble in 250 mL or less of aqueous media over the pH range of 1.7.5. This volume of 250 mL is derived from typical bioequivalence studies that prescribe administration of a drug product to fasting human volunteers with a glass of water.

Q This is a question regarding the interpretation of the Acceptance Table 2 for extended-release products. If the specification is 15�25% for one of the time points, and the table says that �none is more than 10% of labeled content,� how do we calculate the wider range allowed?
A The USP General Chapter <711> Dissolution gives acceptance criteria that may be applied where the monograph does not have a more specific acceptance table. At the L2 level of the test (12 units tested), the criteria refer to the number of results that may fall outside of limits defined by 10% of labeled content. The text explaining the table states that those limits �are expressed in terms of the percentage of labeled amount.� This means that for a monograph specification of 15�25% at a sampling time, the expanded range would be 5�35% (i.e., [15 � 10]% to [25 + 10]%) for samples taken at that time.