dx.doi.org/10.14227/DT150308P45

Question and Answer Section - August 2008

William Brown and Margareth Marques
The following questions have been submitted by readers of Dissolution Technologies. Margareth Marques, Ph. D. and Will Brown, United States Phamacopeia, authored responses to each of the questions.
*Note: These are opinions and interpretations of the authors, and are not necessarily the official viewpoints of the USP

Email for correspondence: web@usp.org

Q The procedure for qualifying USP Apparatus 3 requires two sampling times. It seems that the procedure is intended to ensure that the apparatus is suitable for testing extended-release products. I use it for testing an immediate-release product that only needs to be sampled at one time. How should the qualification of USP Apparatus 3 (reciprocating cylinder) be done if the equipment is not going to be used to evaluate modified dosage forms?
A The USP General Chapter <711> Dissolution states that the performance verification of the USP Apparatus 3 should be done by using the USP Chlorpheniramine Maleate Extended-Release Tablets RS. The detailed procedure on how to carry this verification can be found at http://www.usp.org/pdf/EN/referenceStandards/chlorpheniramineMaleate.pdf. This procedure should be followed regardless of the type of samples that are going to be tested in the apparatus.

Q Can sinkers be used with tablets, or is their use exclusively for capsules?
A Please refer to the USP General Chapter <711> Dissolution, under Apparatus, Apparatus 2 (Paddle Apparatus). Sinkers can be used with any dosage form that floats in the dissolution medium. This general chapter states that a small non-reactive wire helix may be attached to the dosage form to keep it at the bottom of the vessel. An alternative sinker device is described in Figure 2a in this general chapter. This sinker is the one mentioned in the Japanese Pharmacopoeia, and it was incorporated in the USP general chapter when this chapter was harmonized with the Japanese Pharmacopoeia and European Pharmacopoeia. Other validated sinker devices may be used. The USP General Chapter <1092> The Dissolution Procedure: Development and Validation gives more details on how to prepare the wire helix sinker. Sinkers have been used to avoid sticking of film-coated tablets to the vessel walls.

Q What is the difference between degassing and deaeration?
A According to John Burmicz in the Technical Note: Why Is Vibration an Issue for Dissolution Testing?, published in Dissolution Technol. 2008, 15 (1), 29�31, "degassing is a technique whereby dissolved gasses with a high solubility product can be displaced by a gas with a lower solubility product. Typically, helium can be used to displace both nitrogen and oxygen in aqueous media. Deaeration, on the other hand, is a method by which we physically remove the dissolved gasses from the medium."

Q What should be the minimum hardness of the water to be used in the preparation of dissolution medium?
A The water used to prepare the dissolution medium should meet the requirements of the USP monograph for Purified Water.

Q We perform a dissolution profile from 12 tablets. Samples are taken at 5, 10, 20, 30, and 45 min with medium replacement. The monograph requires the test to be evaluated at the 30-min sampling. Is it correct to report the results for the first six tablets from the 30-min sampling? We would make sure that the results for the second set of six tablet were similar to the first.
A With twelve units tested, you should not disregard any data collected. Therefore, the first six could be compared to the S1 criteria in Acceptance Table 1 (General Chapter <711> Dissolution), but you will need to compare the entire set of data to the S2 criteria as well. You should designate in advance the set of tablets that you will consider against the S1 criteria and document that choice. Because the monograph test only includes one test time point, the procedure that you describe is strictly interpreted as an alternative. A discussion of alternative methods is given in the USP General Notices under Tests and Assays.

Q In USP <711> Dissolution, Acceptance Table 1 gives no upper limit for dissolution results. When we are testing the dissolution for an immediate- release product, what should be the maximum value found? According to USP content uniformity, the maximum value should be greater than 100% but less than 115% of label claim. Is this not so?
A General Chapter <711> Dissolution has no upper limit specified for dissolution results summarized in Acceptance Table 1. The emphasis for the performance testing of immediate-release products is on products that do not release an adequate amount. Confirmation of the amount of drug per dosage unit is given by the assay and by <905> Uniformity of Dosage Units. Please note that for <905>, the criteria have changed, and now the calculation of Acceptance Value (AV) is required. Changes to the Uniformity of Dosage Units general chapter are explained by an entry on the USP website under Compendial Notices. The explanatory note can be accessed at http://www.usp.org/USPNF/notices/generalChapter905.html.