dx.doi.org/10.14227/DT150408P35

Question and Answer Section - November 2008

William Brown and Margareth Marques
The following questions have been submitted by readers of Dissolution Technologies. Margareth Marques, Ph. D. and Will Brown, United States Phamacopeia, authored responses to each of the questions.
*Note: These are opinions and interpretations of the authors, and are not necessarily the official viewpoints of the USP

Email for correspondence: web@usp.org

Q We have the following situation: a product releases about 90% of label claim in 7 min in pH 6.8 phosphate buffer, about 40% of label claim in 60 min in 0.1 N hydrochloric acid, and about 50% of label claim in 7 min and about 90% in 60 min in pH 4.6 acetate buffer. Which of these media is the most appropriate for this formulation?
A The discrimination power of the dissolution test needs to be considered when selecting the most appropriate dissolution medium. Fast dissolution is not discriminative enough and is appropriate only for drug substances that are classified as BCS I (highly soluble and highly permeable) when formulated in immediate-release dosage forms. Another aspect to be considered is that the dissolution test should be run in a feasible amount of time. Based on the results obtained with 0.1 N HCl, it looks like it will take some time to release 70�80% of the label claim. In this situation, other parameters such as medium volume and agitation speed could be modified to obtain a more appropriate dissolution profile. The results obtained with pH 4.6 acetate buffer look promising, but it is necessary to verify which medium, hydrochloric acid or acetate buffer, is more discriminative for the critical parameters of this formulation and manufacturing process.

Q Regarding the performance verification of USP Apparatus 1 and 2 using the USP Salicylic Acid Tablets RS, what is the reason to further dilute the sample solution? Why can�t the tablet be dropped and a direct reading obtained?
A The procedure to verify the performance of USP Apparatus 1 and 2 using the USP Salicylic Acid RS is available at: http://www.usp.org/referenceStandards/useAndStorage/calibrators.html The procedure states, "Measure the amount of salicylic acid dissolved from filtered portions of the sample aliquots withdrawn and suitably diluted with Dissolution Medium." This means that it may be necessary to measure the absorbance of samples that have been diluted so that the absorbance is within the recommended range of 0.2�0.8 absorbance units. In any method, for the diluted sample to be compat- ible with the quantitative procedure being used, it must be within the linear range of the analytical technique used. If the absorbance of the undiluted sample aliquot is acceptable, then no dilution is needed.

Q When evaluating dissolution of capsules, must the rotation speed be 100 rpm or is it possible to use other speeds such as 50 or 75 rpm for baskets?
A The agitation speed of the basket is selected by taking into account how the particles are being dispersed in the dissolution medium and the discrimination power of the test regarding the critical parameters for the formulation under evaluation. It is not advisable to use speeds above 100 rpm unless justified to give results that reflect in vivo performance or to provide better discrimination without adversely affecting method reproducibility. The decision to use a particular rotation speed (or other test conditions) will be based on the evaluation of dissolution profiles obtained under various conditions.

Q Please describe the calculation of the amount released when using an open-loop USP Apparatus 4 (flow-through cell).
A If an online UV spectrophotometer is being used, the sample absorbance is measured at regular intervals, the concentration calculated, and the dissolution rate determined as the multiplication product of the concentration and the flow rate. This release rate at each sampling time can be reported, or the cumulative amounts dissolved can be estimated for each sampling interval. The cumulative amount dissolved can be estimated from the average dissolution rate at the beginning and ending time points multiplied by the elapsed time (trapezoid rule). The calculations can be done by hand or by using a spreadsheet.
Alternatively, the cumulative amount dissolved can be measured exactly using an off-line spectrophotometer. The drug concentration in the fraction collected over the entire interval is the most accurate measure of the amount of drug released. Furthermore, the volume collected gives a better confirmation of the actual cell flow rate during the interval.

Q Could the USP Salicylic Acid Tablets RS be used to qualify the USP Apparatus 4 (flow-through cell)?
A Up to now, the USP Salicylic Acid Tablets RS have not been fully validated to verify the performance of USP Apparatus 4. There are papers in the literature, such as Kauffman, J. S. Qualification and Validation of USP Apparatus 4. Dissolution Technol. 2005, 12 (3), 41�43, that have evaluated the use of this USP Reference Standard in the qualification and performance evaluation of this apparatus, but this use it is not official or compendial.

Q What could be the volume of dissolution medium for the evaluation of drug release from buccal patches or tablets?
A The minimum volume of dissolution medium to be used in the evaluation of drug release from any dosage form is defined by taking into account the sink condition of the drug substance. As the volume of fluid in the buccal pouch is considerably less than the typical volumes encountered in dissolution testing of tablets and capsules, a system that can provide dissolution results based on similar volumes might be useful in the evaluation of the performance of buccal patches or tablets during dosage form development. Several designs for small-volume dissolution test apparatus are available. The most important concern for any dissolution test is that the set conditions are sensitive to product changes that may adversely affect the therapeutic effect.

Q Is there a minimum amount of drug to be released when defining the Q value for a particular dosage form?
A No, there is no minimum amount to be released. During the development of an immediate-release product, a target Q of about 70�80% released in 30�45 minutes is commonly used until more information about the product becomes available. The final definition of the Q value is going to be made after considering the dissolution profiles of all batches (pilot, manufacturing, stability) and the probability of passing the S1, S2, S3 criteria defined in the USP General Chapter <711> Dissolution. More information can be found the in following papers: (1) Dumont, M. L., Berry, M. R., Nickerson, B. Probability of passing dissolution acceptance criteria for an immediate release tablet. J. Pharm. Biom. Anal. 2007, 44, 79�84; (2) Hauck, W., et al. Oral dosage form performance tests: new dissolution approaches. Pharm. Res. 2005, 22 (2), 182�187; and (3) Cholayudth, P. Using the Bergum method and MS Excel to determine the probability of passing USP dissolution test. Pharm. Technol. 2006, 30 (1), 88�94.

Q What is the difference between artificial gastric fluid and simulated gastric fluid?
A Various formulas can be found for dissolution media that approximate gastric fluid. Where a reference is made to USP, the preparation of simulated gastric fluid can be found under Test solutions, in the USP section Reagents, Indicators, and Solutions.

Q What are the reasons for quantifying the amount of drug release at the isosbestic point and not at the wavelength of maximum absorption?
A Isosbestic point is a specific wavelength at which two or more chemical species have the same concentration-based absorptivity. When superimposing the absorption spectra of these two species at equivalent concentrations (molar or mg/mL), the isosbestic point corresponds to the wavelength at which these spectra cross each other. An example is shown above in the spectrum of bromocresol green (source: Wikipedia). Isosbestic points are useful in quantifying the amount of drug released when the dissolved drug undergoes degradation into a defined product, such as aspirin, which hydrolyzes to salicylic acid (see Aspirin Tablets monograph in USP 31).

Q Is it acceptable for a tablet or capsule to float during the dissolution test? Must the floating dosage form be weighed down with a sinker to keep it at the bottom of the vessel? Must the dosage form sink to the bottom of the vessel? Must a sinker be used?
A The dosage form should lie at the bottom of the vessel during dissolution testing with the paddle apparatus. The various regions in the vessel are associated with different characteristic hydrodynamics, and consistency of results is more likely when the sample remains in place. A floating sample may move into areas of high or low flow within the medium and may encounter the moving paddle, leading to a mechanical interaction that may impair the integrity of the dosage form. Any of these events may alter the dissolution rate of the individual unit under test and increase the variability of the results. The use of a sinker is often included in a method description and occasionally part of the monograph procedure. The sinker design has a great influence in the dissolution profile of the dosage form. A number of different types of sinkers are available, and when sinkers are used, the particular design employed should be well described in the method documentation.

Q I am testing an extended-release product using the criteria given in the dissolution general chapter. The procedure calls for sampling at three time points, 1, 4, and 8 h. For the first six tablets tested, the results met the criteria in Acceptance Table 2 at the 1- and 8-hour time points but did not meet the criteria at the 4-hour time. When we repeated the test with the next six tablets, the results were acceptable at each time point. Did the product pass the test at L1 for some time points and at L2 for the remaining time point?
A he performance of the product is measured over all time points. Under the Interpretation section of the general chapter, the instruction is to continue testing through the three levels unless results conform at either L1 or L2. In the case given, the results did not conform at L1 (six units tested) and you continued to L2 testing (six additional units for a total of twelve). The procedure calls for sampling at each time point, which you did. Having results for all time points, you compared with the acceptance table. All results for L2 testing are used if testing needs to proceed to that level.