dx.doi.org/10.14227/DT200113P60

Question and Answer Section - February 2013

William Brown and Margareth Marques
The following questions have been submitted by readers of Dissolution Technologies. Margareth Marques, Ph. D. and Will Brown, United States Phamacopeia, authored responses to each of the questions.
*Note: These are opinions and interpretations of the authors, and are not necessarily the official viewpoints of the USP

Email for correspondence: web@usp.org

Q Which solvents can be employed to prepare the standard solution to be used in the quantitative procedure in dissolution testing? Is it mandatory to use the dissolution medium?
A There is no general answer to this question. The conditions will be defined in a case-by-case approach depending on the quantitative procedure characteristics. Once the dissolution sample is filtered, provided the filtration step was properly validated, you can do whatever is necessary to have reliable quantitative results. Depending on the case, additional dilutions can be made using any solvent, such as the mobile phase, or a derivatization reaction can be done to generate a compound for which a specific detector has higher sensitivity, or a chemical reaction can be done to remove an interference, and so forth.

Q During a multimedia dissolution study of our formulation, we found an irregular peak shape for the active ingredient in 0.1 N hydrochloric acid because the dissolution medium was not compatible with chromatographic conditions. Consequently, the criteria for peak symmetry, theoretical plates, and relative standard deviation could not be met. A second dilution in water was made and injected into the chromatograph, and a better peak shape was obtained. Can other diluents, other than the dissolution medium, be used to prepare the samples in the dissolution testing?
A See the answer to the previous question.

Q Is it necessary to run a dissolution test for an effervescent tablet?
A If using the tablet according to the label instructions results in a solution, the dissolution test is not needed. Other quality parameters, such as completeness of solution, may need to be evaluated. If the tablet produces a suspension, then a dissolution test needs to be developed.

Q Should the evaluation of accuracy/recovery in the validation of a drug release test for a transdermal system be done at 50% of label claim or 50% of the lowest amount of drug released during the experiment?
A When doing an accuracy/recovery evaluation of dissolution procedures (regardless of the dosage form type or dissolution apparatus), most companies select at least three concentrations from the linearity range of the method. For the highest concentration, the upper limit is defined by the allowable range from uniformity of dosage units. Most transdermal systems (TDS) have two label claims, one is the total drug load present in the TDS (TDS have a labeled overage to provide the appropriate drug delivery over the treatment period) and a second label claim for delivered dosing profile of drug released. The value used in all calculations for TDS will be based on the dose released from the TDS.

Q The linearity range was defined for the validation of the drug release test for a transdermal system (TDS), but what is the lowest concentration level to be used for the accuracy evaluation?
A The range for accuracy comes from the range for linearity. At least three concentrations from the linearity range should be used in the accuracy evaluation. If the product has different doses, all of them must be considered when establishing the linearity range and defining the concentrations to be used to evaluate accuracy.

Q We are thinking of buying some new vessels for our dissolution apparatus. We are trying to decide what type of vessels to buy, round-bottom vessels (that we are using now) or Peak vessels. Since we have no experience with Peak vessels, we would like to know their advantages and disadvantages compared with round-bottom vessels. Is there any regulation that says we should not use Peak vessels in some cases (e.g., dissolution profile comparison)? If we buy Peak vessels, should we revalidate all of our dissolution methods? What is meant when USP vessels are specifically mentioned in some literature articles? Is this a round-bottom vessel or some other type, or is does this mean any vessel that complies with USP requirements?
A Peak vessels were developed in an attempt to minimize perceived problems in the dissolution tests due to the presence of a cone of granulate or particles on the bottom of the round-bottom vessels. Peak vessels should be used with caution because the issue of coning can be solved in most cases by increasing the rotation of the paddle. Peak vessels are not compendial apparatus. They are not included in the USP or European Pharmacopoeiaor Japanese Pharmacopoeiageneral chapters on dissolution. They are not standardized. They can only be used in very special situations with very good scientific justification. Keep in mind that cone formation can be reduced or eliminated by increasing the paddle rotation. If you have dissolution methods developed with round-bottom vessels, you cannot change these methods to use Peak vessels. Peak vessels have a very specific application and can only be used with very good justification. USP vessels are the vessels described in the USP General Chapter <711> Dissolution. They are round bottom, and they can have volumes of 1 L, 2 L, and 4 L.

Q What should be the paddle and basket heights when using a Peak vessel?
A There are only two USP monographs that use Peak vessels, praziquantel tablets (veterinary use) and galantamine tablets, and both use the paddle at 2.5 cm.

Q During the evaluation of dissolution results in a stability study for an immediate-release dosage form, is it necessary to go to Stage 2 when only one unit is below the acceptance criterion or when more than one unit fails Stage 1 (S1)?
A The dissolution test is performed in three stages, if needed. The test can be concluded when the results conform at any stage. The S1 criteria require that results are not less than Q + 5%. If that condition is not true, then the testing proceeds to the next stage. The USP standard applies over the shelf life of the product.

Q Which reference solution concentration should be used during the performance verification testing for USP Apparatus 1 and 2?
A The USP General Chapter <851> Spectrophotometry and Light Scatteringrecommends that the standard solution have a concentration within about 10% of the test solution. When the standard and test solution concentrations are in a linear range with respect to the absorbance (Beer–Lambert relationship obeyed), the differences can be greater. The USP Dissolution Toolkit, available at http://www.usp.org/sites/default/files/usp_pdf/EN/dissolutionProcedureToolkit201003.pdf, recommends that a standard solution of 0.01 mg/ mL prednisone be used. This corresponds to 50% of the 10-mg prednisone Reference Standard Tablet (RS tablet) dissolved in 500 mL of medium. The certificate provided with the RS tablet just recommends the use of a standard solution of known concentration. The response of prednisone solutions in water is established for concentrations relevant to the PVT. However, the performance of the spectrophotometer in your lab is a variable that should be evaluated, so you should check that you are getting a linear response in the concen-tration range of interest. Once you establish linearity, you should choose a concentration appropriate to the analysis of the prednisone solutions obtained in your PVT experiments. Standardization of procedure is viewed favorably, and documentation of your decisions is part of good practices.

Q We are developing a dissolution method for a combination product containing cefixime and azithromycin. The USP monographs give individual dissolution media for each drug. How is a dissolution medium selected for a combination product?
A You need to verify experimentally the best dissolution conditions for this type of product. One of the major players in helping select the conditions is the solubility of each drug substance in dissolution media within the physiological pH range at 37 °C. For combination products, there are no specific rules. You can have the same dissolution conditions for both drug substances, you can have the same dissolution conditions but different acceptance criteria, or you can even have two separate dissolution procedures for each one of the active ingredients.

Q We are developing suppositories containing paracetamol, and we found that a high speed (175– 200 rpm) should be used to release approximately 90–95% of the active ingredient in 1 h. Is this high speed acceptable?
A The dissolution test is a way to evaluate the performance, in vitro, of the dosage form. The main objective of the dissolution test is not to release the total amount of drug present in the dosage form. The dissolution test should be discriminative for the critical quality attributes associated with the product. If there is any deviation in any of these critical attributes, the dissolution profiles should show some differences. At the early stages of development of a dissolution test, the target should be to release around 70–80% of the product label claim. The final acceptance criterion will be selected as the point in the dissolution profile determined to be the most discriminative. Often a slower dissolution profile may produce a more discriminative dissolution test.

Q I am preparing a film containing sumatriptan. How should I select the composition of the dissolution medium?
A The composition of the dissolution medium depends on the physiological properties of the route of administration of the product (oral, topical, injection, etc.), the region in the body where most of the absorption will occur (mouth, skin, stomach, etc.), and the solubility of the active ingredient in the conditions (temperature, pH, etc.) of the route and region of absorption.

Q We are running a dissolution test for gelatin capsules, and we need to add pancreatin to the dissolution medium because the capsules are showing cross-linking. We used the information on the product label to calculate the amount of enzyme needed, and we added the calculated amount to each vessel. Is this the right way of defining the correct quantity?
A The information on the product label is just the manu-facturer specification, and it is not the enzymatic activity of the particular batch of pancreatin that is going to be used in your lab. You have two options: (1) contact the reagent supplier and ask for the enzymatic activity of the particular lot you plan to use and the method that was used to determine its activity (it should be the protease activity determined according to the USP monograph for Pancreatin), or (2) determine the protease activity in your lab using the procedure in the USP monograph for Pancreatin. The medium containing the enzyme should be prepared in bulk rather than by adding the enzyme to each vessel. Preparing the enzyme-containing medium in bulk will reduce the variability in the results by homogenizing the medium.