dx.doi.org/10.14227/DT200213p61
dx.doi.org/10.14227/DT200213P61

Question and Answer Section - May 2013

William Brown and Margareth Marques
The following questions have been submitted by readers of Dissolution Technologies. Margareth Marques, Ph. D. and Will Brown, United States Phamacopeia, authored responses to each of the questions.
*Note: These are opinions and interpretations of the authors, and are not necessarily the official viewpoints of the USP

Email for correspondence: web@usp.org

Q In the Dissolution test in the USP monograph for Divalproex Sodium Extended-Release Tablets, it says to filter the sample solution through a 20-µm filter. Why is this filter used if the sample for HPLC analysis needs to be filtered through a 0.45-µm filter?
A Filtration is one of the most important steps in a dissolution test. It stops the dissolution process by removing all particulates in suspension, including undissolved particles of drug substance. The filter used in dissolution is selected in a case-by-case approach. The selection of the filter pore size and filter material will depend on the formulation composition. The filter must be evaluated for possible adsorption of the active ingredient and for possible interference of leachables, extractables, or both in the quantitative procedure. The verification of the filter used will demonstrate that it is adequate to produce a filtrate that can be used in the HPLC analysis. See more information in the USPGeneral Chapter <1092> The Dissolution Procedure: Development and Validation.

Q Most of the dissolution tests use 900 mL of medium. How was this volume defined? Is there any correlation with the volume of gastric fluid in humans?
A The volume of dissolution medium to be used is defined considering "sink condition." The solubility of the drug substance is quantitatively determined in several dissolution media within the physiological pH range at 37 °C. Using this value, the volume of dissolution medium necessary to obtain a saturated solution of the highest dose of the product to be marketed is calculated. Sink condition is considered as at least 3 times this volume. Some companies work with 5 times or 10 times this value. There are some instances where the dissolution test is more discriminative if sink condition is not followed. The volume of dissolution medium has no relationship with the volume of gastric fluids in humans. According to information available in the literature, a realistic volume to simulate the total fluid available in the stomach in the fasted state is in the range of 250-300 mL.

Q I have been assigned to develop a dissolution test for an extended-release capsule product for which I need to change the medium from acid to buffer after 4 h. My product contains granules with different particle sizes, from about 150 to 500 µm. With the paddle apparatus after 4 h in acid, all these granules are floating in the medium. How can I make the change to the other medium without losing granules?
A USP Apparatus 3 and 4 may be good options in a case such as this one. The appropriate screen mesh can be used in the bottom and top of the reciprocating cylinder in Apparatus 3, and it will retain the granules inside the cylinder when the rack is moved from the acid medium row to the next row of tubes containing the buffer medium. Apparatus 4 will contain the granules inside the cell. When the dissolution medium leaves the cell, it passes through a filter retaining all the granules inside the cell. If you are using USP Apparatus 2, the option is to put a filter or mesh screen at the tip of the cannula used to remove the medium by vacuum. After the removal of the acid medium, the buffer medium can be added through the same cannula washing all the granules back to the vessel.

Q I am working with a gastroretentive drug delivery system for a fixed combination product. Drug A has absorbance in 0.1 N HCl while drug B does not. Can I use two separate media for the fixed combination product, 0.1 N HCl for drug A and water for drug B?
A The selection of the dissolution medium is based on the solubility of the drug substance and on the release mechanism of the dosage form. The dissolution medium should be biorelevant, which means that it should have a composition or pH similar to the in vivo region that the dosage form will encounter during in vivo dissolution and that absorption of the drug substance will occur. Gastroretentive dosage forms have longer gastric residence times than other solid oral dosage forms. This approach is used to increase the time for dissolution of the drug in the stomach. Therefore, the composition of the medium for this type of dosage form should have a pH close to the pH of the stomach, and this is not the case with water. Water would only be used if it is a more discriminative medium.

There is no standard procedure for dissolution tests for combination products. The same conditions can be applied to all drug substances present in the dosage form, some of the conditions can be different, or only the acceptance criterion is different. At the extreme, you can have completely different dissolution tests for each drug substance in the fixed combination product. The selection of the conditions is done in a case-by-case approach.

Q We are developing a gastroretentive dosage form for a combination of antidiabetic drugs containing metformin. It is reported that metformin has absorbance in 0.1 N HCl, but we do not see any peaks in the absorbance spectrum. Can we consider phosphate buffer as dissolution medium for both drug substances? Should we use different dissolution media for both drugs?
A See previous question.

Q We are developing a floating tablet, and it remains floating in the dissolution vessel during the entire test. Is it mandatory that the tablet should remain on the bottom of the dissolution vessel?
A Your tablet needs to have good contact with the dissolution medium. You can use USP Apparatus 1 (basket), USP Apparatus 2 (paddle) with sinkers, USP Apparatus 3 (reciprocating cylinder), or USP Apparatus 4 (flow-through cell). Apparatus 3 may be a good option in this case because of its higher turbulence when compared with Apparatus 1 and 2. The selection should be done in a case-by-case approach. You need to select the dissolution conditions that are the most discriminating for each product.

Q I am preparing a dissolution medium, and when I mix the buffer with a 2% sodium lauryl sulfate solution, I get a cloudy solution. Is this because of temperature? What could be done?
A The phenomenon you are facing has nothing to do with temperature. Probably you are using the potassium salt of the reagent in the preparation of the buffer. Sodium lauryl sulfate is not compatible with potassium salts. Try using the sodium salt of the reagent.

Q We are developing a 200-mg tablet containing cefpodoxime. Our product is bioequivalent to the reference product, but it fails the dissolution test in three media (pH 1.2, pH 4.6, and pH 6.8), and the results for f2dissolution profile comparison are less than 50. What can we do for the biowaiver request for cefpodoxime 100-mg tablet?
A Before you start developing a dissolution test, you need two very important pieces of information:

(1) Solubility of the drug substance at 37 °C in various dissolution media within the physiological pH range. The solubility needs to be determined with the actual drug substance that will be used in manufacturing the dosage form. This information will help in selecting the dissolution medium and defining the biopharmaceutics classification (BCS) of the drug substance. Also, it is important to verify any information available in the literature about a possible BCS classification for the drug. According to some papers, cefpodoxime proxetil is a BCS Class 4, which means low solubility and low permeability.

(2) Composition and release mechanism of the dosage form. This information will also help in defining the dissolution conditions. Cefpodoxime has very low solubility in aqueous media. Each company may formulate the product in a different way to overcome or minimize this problem. Consequently, the in vitro performance will be formulation dependent. The objective is a dissolution test that is discriminative for the formulation. In the case of a low solubility drug substance, the pH range for the various dissolution profiles will be defined around the pKaor pKb of the drug substance. A dissolution medium with a pH value outside of the physiological range requires proper justification.

Biowaivers for lower doses of a product that contains a low solubility drug substance may be possible only if the formulation is proportional to the higher doses and the release mechanism and manufacturing processes are the same.

Q I am working on a project to enhance the solubility of a drug that is insoluble in water. I am uncertain on how to develop the dissolution test; I would like to use phosphate buffer as dissolution medium, but the drug is not soluble in this buffer.
A For the drug to be absorbed, it must be in solution. If the drug is insoluble in water, some formulation strategies should be used to increase its solubility in aqueous solvents. Some of these strategies exploit amorphous solid dispersions, micronize the drug, use surfactants or buffering agents in the formulation, use excipients such as cyclodextrins that create a micro-environment where the drug has higher solubility, use selfemulsifying systems, and so forth. Each formulation will require a different dissolution test depending on the strategy used. If the formulation contains surfactants, the dissolution test will probably be performed using simple buffers or acid solutions. Whatever dissolution conditions are chosen, they should be discriminative for the critical quality attributes such as particle size of the drug substance.

Q Is a dissolution test needed for the active substance as a single component in a capsule dosage form, or would a disintegration test be adequate?
A If the drug substance is a solid, it will still need to dissolve once the capsule shell opens. Disintegration will show that the capsule shell has disintegrated but not that the drug substance has gone into solution. You will first need to develop a suitable dissolution test. Then, depending on the characteristics of your product (if it is BCS Class 1 or 3) and if it can be shown that disintegration correlates with dissolution, you can replace the dissolution test with a discriminative disintegration test. Keep in mind that all the decisions in dissolution/disintegration method development should be justified with data.

Q When transferring a dissolution method to another lab, should we run the dissolution test with 6 or 12 units? Can we use f2as the acceptance criterion for this method transfer?
A There are no official rules for the number of units to be used in dissolution method transfer. It is up to the lab to select the appropriate number of units in each case. There are several approaches that can be used in analytical method transfer (see USPGeneral Chapters <1092> The Dissolution Procedure: Development and Validationand <1224> Transfer of Analytical Procedures). For the transfer of dissolution methods, the recommendation is that each lab should perform a dissolution profile at least in duplicate and evaluate the individual values, mean, and standard deviation at each time point in each dissolution profile. All labs should use the same batch of product. It is up to the labs to define the acceptance criteria for the method transfer, and it may be product dependent.

Q How can I define the time for the dissolution test for ibandronate sodium tablets? Considering the characteristics of this drug substance, is 30 min appropriate? Similar molecules such as alendronate sodium and risedronate sodium have times of 15 and 30 min, respectively.
A The time is selected from the dissolution profiles obtained with all batches tested (pilot batches, manufacturing batches, bio batch, pilot batches with deviations in the critical attributes, and the batches under stability studies). You are going to select the time and tolerance (amount dissolved) so that the dissolution method is the most discriminative for the critical attributes of a particular product.

Q The certificate for a particular lot of USP Prednisone Tablets RS used in the performance verification of dissolution apparatus stated, "measure the absorbance at 242 nm (the approximate wavelength of maximum absorbance)." However, I found that the maximum absorbance in my equipment was 244 nm. The certificate for a different lot of USP Prednisone Tablets RS stated only, "measure the absorbance at 242 nm." What is the appropriate procedure? Should I use the wavelength of maximum absorbance?
A The wavelength accuracy of spectrophotometers depends on the conditions of each individual instrument. That is why the wavelength of maximum absorbance for a particular solution may differ among equipment. The best practice is to use the absorbance maximum obtained with the instrument that is going to be used to perform the analysis. USP has found that for prednisone solutions in water, the maximum will be at about 242 nm. Your observed maximum confirms these statements. Typically, if the observed wavelength is more than 1 nm from a standard value, the wavelength accuracy of the instrument should be checked. A number of standard approaches for this evaluation are given in the USPGeneral Chapter <851> Spectrophotometry and Light Scattering.

Q The USP monograph for Orlistat Capsules calls for the use of a wire coil sinker. We were not able to find a vendor for this type of sinker. Could you please give the specification for this sinker or suggest a possible supplier?
A The instructions on how to prepare this type of sinker are given in the USPGeneral Chapter <1092> The Dissolution Procedure: Development and Validation, under Sinkers. Additional information on the selection of sinkers for capsules can be found in the USPGeneral Chapter <1094> Capsules—Dissolution Testing and Related Quality Attributes (published in Pharmacopeial Forum39(3), available free of charge at www.usppf.com). Sinkers play a very important role in dissolution, and the design and size used should be selected in a case-by-case approach.

Q If a USP monograph has multiple dissolution tests, should we perform all of them on our product or can we choose one of them?
A The dissolution tests in any USP monograph are the tests approved by FDA for the products to be marketed in the U.S., and they may not be appropriate for products available in other regions. For products approved for the U.S. market, the appropriate dissolution test should be part of the labeling. Products approved by the FDA are listed in the Orange Book. The Orange Book is available free of charge at http://www.accessdata.fda.gov/scripts/cder/ob/default.cfm.

For products that have not been approved by the FDA, the first step in evaluating the suitability of any of the dissolution tests in USP monographs is to run a dissolution profile with the conditions stated in the monograph. A test procedure may be suitable for your formulation if it gives results that are similar to the times and tolerances stated. It is important to remember that a suitable dissolution test will have sensitivity to changes in manufacture and on stability. Once a promising procedure is identified, you need to do a complete validation, including validating the filter and deaeration procedure (see USPGeneral Chapter <1092> The Dissolution Procedure: Development and Validation). If none of the tests is suitable for your product, you need to develop a new test discriminative for your formulation.

Q In the dissolution test of a particular USP monograph, the standard solution is prepared at a specific concentration in dissolution medium. The drug substance has very low solubility in medium. How can we prepare this solution?
A The USPGeneral Chapter <1092> The Dissolution Procedure: Development and Validationrecommends that when the compound has low solubility in dissolution medium, a small amount of organic solvent, in most cases not more than 5% of the final volume of the first dilution, may be used to help dissolving the compound. Typically, solutions within that concentration limit do not pose an interference to the analysis. Interference to the analysis should be evaluated, especially if a higher amount of organic solvent is needed.

Q Is it advisable to dilute the aliquot of a dissolution sample or is it better to inject the aliquot withdrawn directly into the instrument?
A It depends on the linearity of the quantitative method that is used to determine the amount of drug released. The final concentration of the sample solution should be within the linear range of the quantitative method. Once the sample solution is filtered, provided the filtration step was properly validated, it can be treated in any suitable way to give reproducible results.

Q Why does the USP monograph for Glimepiride Tablets have three different dissolutions tests?
A Glimepiride is practically insoluble in aqueous solvents. Because the drug needs to be in solution to be absorbed, each company is going to use a different formulation- manufacturing process strategy to increase the solubility of the drug substance. Therefore, the dissolution test will be formulation dependent. Each one of the three tests is specific for a particular product approved by FDA for the U.S. market, and they may not be appropriate for other formulations.

Q I would like to know if there are any USP monographs or FDA approved products that use USP Apparatus 4.
A The only USP monograph that calls for the use of USP Apparatus 4 is Rufinamide Tablets. You can find the dissolution conditions recommended by the FDA Office of Generic Drugs at the web site http://www.accessdata.fda.gov/scripts/cder/dissolution/index.cfm There you can find the following products approved with a dissolution/ drug release test using USP Apparatus 4: betamethasone acetate and betamethasone sodium phosphate injectable suspension, methylprednisolone acetate injectable suspension, and rufinamide tablets.