dx.doi.org/10.14227/DT200413P51

Question and Answer Section — November 2013

William Brown and Margareth Marques
The following questions have been submitted by readers of Dissolution Technologies. Margareth Marques, Ph. D. and Will Brown, United States Phamacopeia, authored responses to each of the questions.
*Note: These are opinions and interpretations of the authors, and are not necessarily the official viewpoints of the USP

Email for correspondence: web@usp.org

Q In the performance verification (PVT) testing with prednisone tablets, our mean values satisfy the specification given in Table 1 of the certificate that comes with the tablets ( prednisonelotq1l136.pdf ). We would like to know if the individual values obtained with each vessel must satisfy this specification as well. The individual values that we obtained are 56, 55, 67, 63, 57, 63, and 61, and the geometric mean (GM) is 60.
A From your data, we assume that you are testing the basket apparatus and that you are performing the first stage of the two-stage test with seven vessels. For this condition, the acceptance criterion is a GM between 58% and 72%, and the limit for acceptable %CV (coefficient of variation) is 9.2%. Your data give a GM of 60% and a %CV of 7.3%. Both of these values are within the acceptable ranges. You indicate that individual values lie outside the acceptable range for the GM, and that is correct. The test is set up to limit the GM and not individual values. Therefore, your concern is unwarranted. Instead of establishing a limit to the acceptable range of individual values, the distribution of the values is controlled by the %CV. Again in your case, the %CV that you observed is within the acceptable range.

Q We are carrying out a dissolution test according to the USP monograph. However, we are having slightly lower dissolution values by following the USP dissolution test, and with peak vessels, the dissolution is improving. Can we use peak vessels in this case?
A Peak vessels are not compendial apparatus, and they should only be used after you have tried other compendial conditions. They can only be used with very good justification. In most cases, increasing the rotation speed of the paddles gives results similar to those obtained with peak vessels. Be careful with the evaluation of your results. If the dissolution test was appropriate for your product in the past and now you are seeing differences with your dissolution results, this may be an indication that something is different with your product and you need to investigate.

Q In one of the USP monographs for orally disintegrating tablets, there are two disintegration tests with different acceptance criteria, one is NMT 10 s and the other is NMT 30 s. What are the reasons for different acceptance criteria?
A Orally disintegrating tablets can be manufactured by two processes: (1) lyophilization, which produces a very soft and fragile "tablet" that disintegrates in just a few seconds and (2) use of super disintegrants, which produces a harder tablet. We might expect case 1 to produce a product that would disintegrate faster than tablets made according to case 2. This is the reason for the two Disintegration Tests. The procedure for the test is described in the USP General Chapter <701> Disintegration. The acceptance criteria for disintegration of orally disintegrating tablets may be formulation dependent, so you will need to define the appropriate acceptance criteria for your formulation. Typically, orally disintegrating tablets need to have a disintegration and a dissolution test. For the FDA guidance for orally disintegrating tablets, see UCM070578.pdf .

Q Our product is a tablet containing a drug substance that belongs to BCS Class 2. The recommendation in the FDA guidance for establishing the dissolution acceptance criteria is to choose two time points, one at 15 min and the other at a time when the release is equal to or greater than 85%. However, our product contains more than 10% in weight of coating, and at 15 min, we are obtaining only 10% of drug released. In this case, is 15 min reasonable as the first time point? Should we choose the second time point at 30 min (30% released) or 45 min (60% released)?
A The acceptance criteria are selected from the dissolution profiles obtained from all batches evaluated during product development (pilot batches, batches with deviations in the key quality attributes, bio-batch, etc.) including the batches under stability studies. We cannot speak to the FDA expectations. From a compendial point of view, two-point dissolution test procedures are not mandatory for immediate-release products. From your experience, you need to verify the sampling times and criteria that add information and that are the most discriminative for the key quality attributes.

Q In one of the USP monographs, a dissolution test states to deaerate the dissolution medium with helium. How is this done?
A This procedure is done by bubbling helium directly into the dissolution medium for a certain period of time. Helium is very expensive and not readily available everywhere. Other deaeration procedures can be used. A description of a deaeration procedure that uses heat and vacuum is in the USP general chapter. Also, have a look at the paper "Comparison of the Effectiveness of Various Deaeration Techniques" in Dissolution Technol. 2004, 11 (1), available at www.dissolutiontech.com.

Q The FDA guidance recommends the use of 12 units of the dosage form for the comparison of the dissolution profile of the reference and the generic products. To obtain the results from the 12 units, can we run the test three times with four units?
A Most of the dissolution equipment available on the market has six or more vessels. To obtain 12 results, you just need to do two runs of six units with each product. As far as we know, there are no rules on how to obtain the 12 results, but most labs do two runs with six vessels.

Q How is the dissolution sample solution determined? We find that in some tests, the samples withdrawn from the vessel need to be diluted as part of the quantitative procedure.
A The final concentration of the sample solution depends on the linearity of the method and the analytical technique. Sometimes you will need to dilute the sample solutions to be within the linear range of the technique and the detector. Adjust the calculated result based on any dilutions done.

Q What cell path length should be used in the Dissolution Test 2, Buffer Stage, in the USP monograph for Omeprazole Delayed-Release Capsules?
A If the text does not mention the cell path length, it is assumed that a 1-cm cell should be used. The text will mention the cell size only if it is different from 1 cm.

Q We validated a dissolution method for a tablet using USP Apparatus 1. There was a modification in the compression step of this tablet, and the basket apparatus is not appropriate anymore. We changed the method to use paddles with all the other parameters remaining the same. Is it necessary to revalidate the method?
A The change that was made is very critical. The modified method is considered a new method. You will need to demonstrate that the modified method is discriminative and justify the selection of all parameters and acceptance criteria. The modified method requires full validation.

Q We are trying to use Dissolution Test 4 in the USP monograph for Pantoprazole Sodium Delayed- Release Tablets, which specifies the use of sinkers. If we use this test without the sinkers, should we validate the method or just verify it? In our experience, some enteric coatings have adhesive behavior with sinkers and slow the dissolution of the product.
A Sinkers are very critical, and they can have a big impact on the dissolution profile. Depending on their design, they can even prevent the disintegration of the dosage form. The selection of the sinker design is formulation dependent. Each formulation will require a specific type of sinker. Sinkers are used if the dosage form floats or if it sticks to the vessel wall. First you need to verify that your formulation requires the use of sinkers. If it does, you need to select the most appropriate type and size for your formulation. There are several designs and types of sinkers, made of different materials, available on the market. If the procedure in the monograph is changed in any way, you will need to validate the new dissolution test.

Q Is the USP monograph for Isotretionin Capsules applicable for hard or soft capsules? Why there are three different dissolution tests in this monograph?
A The USP monograph is applicable to any Isotretinoin Capsules approved by FDA for the U.S. market regardless of the type of capsule shell (hard or soft, gelatin, starch derivative, or cellulose derivative) and the type of filling (solid, liquid, or semisolid). Isotretinoin is practically insoluble in aqueous solvents. Each company uses a different formulation strategy to minimize or overcome this problem. Consequently, each formulation is going to have its own dissolution test. Each dissolution test in this monograph is specific for a particular product approved by FDA for the U.S. market.

Q If the assay test is done by HPLC or by spectrophotometry and the dissolution test uses the same quantitative method, can it be assumed that the validation of the assay test can be used for the dissolution test?
A No, the sample is treated in very different ways in the assay and dissolution tests, and each test needs to be validated separately. The required analytical range and the influence of excipients will be different for each test. In the case of spectrophotometric methods, there is chance that the blanks may be different for each test. For dissolution, it is especially necessary to check for filter interference.

Q I am designing a new apparatus for the in vitro release of floating drug delivery systems. Do I need to have a procedure for the performance verification (PVT) of this new apparatus?
A You will need detailed drawings of this new apparatus with all the measurements and tolerances and a description of the materials of construction. All documentation needs to be written in such a way that the apparatus can be replicated in any facility. In addition, you need to determine the critical attributes of your apparatus and their impact on the drug release determinations, and specify acceptable tolerances. Confirming that the apparatus conforms to specified dimensions and operational parameters such as flow rate or rotation speed will be an important part of the qualification. Finally, the purpose of the apparatus is to produce a sample representing the in vitro performance of a dosage form. Given that purpose, suitable performance of an apparatus will be demonstrated from acceptable dissolution results using a standardized sample material.

Q We are investigating the solubility of several drug substances under different pH conditions for the development of veterinary pharmaceutical feed additives. The FDA Guidance for Industry Waiver of In Vivo Bioavailability and Bioequivalence Studies for Immediate-Release Solid Oral Dosage Forms Based on a Biopharmaceutics Classification System (BCS) ( ucm070246.pdf ) says to verify the pH of the solution after the addition of the drug substance to a buffer. It is necessary to have constant monitoring of the pH of the solution throughout the solubility determination?
A The pH range may be selected based on the animal species to which the product will be administered. Depending on the animal species, the pH ranges in the BCS guidance may not be applicable. The pH range in the BCS guidance may be suitable for dogs. Papers that discuss this topic are "Solubility Criteria for Veterinary Drugs—Workshop Report" in Pharmacopeial Forum, 39(4) [Jul.-Aug. 2013], "Solubility Criteria for Veterinary Drugs" in Pharmacopeial Forum 38(4) [Jul.-Aug. 2012], and "Veterinary Application of In Vitro Dissolution Data and the Biopharmaceuticals Classification System." All are available at keyissue-standards-veterinary-drugs .

Determine the pH of the solution before adding the drug substance, after adding the drug substance, and at the end of the solubility experiment to verify that the buffer capacity of the selected buffer is adequate. Certain compounds may react with the medium to alter the pH, possibly affecting the solubility of the compound.

Q Please explain the statement "Validation studies should address the variations associated with different profile time points" found in the USP General Chapter <1092> The Dissolution Procedure: Development and Validation.
A When running a dissolution profile, in most cases higher variability in the dissolution results is seen at the initial timepoints. The variability should decrease with time. Additionally, the possibility of interference from the excipients may be different over the dissolution profile as the formulation interacts with the dissolution medium.

Q Why are there three different dissolution tests in the USP monograph for Glimepiride Tablets? How should the correct one be selected for use?
A Glimepiride is practically insoluble in aqueous solvents. To overcome or minimize this problem, each company will formulate the product using a different strategy based on available technology or patent issues, cost analysis, and so forth. The dissolution test for each formulation will reflect those differences. The dissolution tests in USP monographs have been approved by FDA for products marketed in the United States. They may not be appropriate for products sold in other regions. The dissolution tests in the USP monograph can be used as a starting point in the development of dissolution tests for new formulations.

Q We are evaluating the dissolution of esomeprazole delayed-release capsules and are finding evidence that possibly 20-30% is released in the acid-stage testing. Esomeprazole is unstable in acid medium. How can we evaluate any release in the acid medium?
A Acid-resistant formulations should release very little drug in the acid stage. The USP General Chapter <711> Dissolution limits release in the acid stage to not more than 10% of the product label claim. If you are finding high values released in the acid stage, you should discuss your observations with your formulators or product development group. If the drug substance is unstable in acid medium, as is the case of omeprazole, lansoprazole, esomeprazole, and other drug substances from the same family of compounds, you can determine the amount that remains in the dosage form after it has been submitted to the acid-stage test. You can do this using the assay or the quantitative procedure from the uniformity of dose test.

Q Can mini vessels be used in the dissolution of tablets?
A Mini vessels are not compendial apparatus, and the marketplace is not standardized. They should be used only with proper justification such as demonstrating that they bring an advantage to your dissolution method when compared with the compendial apparatus. Since they are not standardized, you need to describe the vessels (dimensions, shape, supplier, and catalog number) to be used in your method.

Q In the USP monograph for Atenolol Tablets, the acceptance criterion for the dissolution test is NLT 80% (Q) in 30 min. The manufacturer's specification is NLT 85% in 30 min. Is this specification in compliance with USP?
A The USP monograph for Atenolol Tablets gives a dissolution tolerance with the Q value of 80%. Acceptance Table 1 in the USP General Chapter <711> Dissolution shows that at Stage 1, testing 6 units, no unit result is less than 85% (Q + 5%). The manufacturer may mean that its specification is set at the S1 level as presented in Acceptance Table 1. That would partially conform with USP. The general chapter describes the dissolution test as having three stages of testing, S1, S2, and S3, with a possible total of 24 units tested. The Q value is used in different ways as the testing proceeds through all the stages. At the S2 level with 12 units tested, Q is the limit for the average result with Q - 15% as the limit for individual results. If the manufacturer simply means that Q is 85% for its atenolol tablets, then it is applying a more stringent requirement than the USP monograph test. That may be done as a control on the product in its hands; however, it is not required for products labeled as USP or claiming to meet USP monograph requirements.

Q In the USP monograph for Oxybutynin Chloride Extended-Release Tablets under Dissolution Test 4 is the recommendation for a specific sinker with its catalog number and a possible supplier for it. The web site for this supplier mentions that sinkers are for capsules, and the USP monograph uses the same sinker for a tablet. Is this correct? Can other types of sinker be used?
A Sinkers can be used with either capsules or tablets if they float or if they stick to the vessel walls (see USP General Chapter <1092> The Dissolution Procedure: Development and Validation). The type of sinker can have a big impact on the dissolution profile; depending on its design, it can even prevent dosage form disintegration. Therefore, the type of sinker used should be carefully evaluated, and sinker selection should be done on a case-by-case basis.

Q There is a discrepancy between the temperature for the deaeration procedure in the USP General Chapter <711> Dissolution (41 °C) and in the certificate that comes with the USP Prednisone Tablets RS (41-45 °C). What are the reasons for this discrepancy? Is there a difference in the deaeration of the medium at a temperature of 41 or 45 °C?
A The underlying principle is that the water solubilities of atmospheric gases decrease with increasing temperature and decreasing pressure. The medium is heated above the test temperature, and then vacuum is applied with filtration and stirring to decrease the dissolved gas content below saturation. Exposed to the atmosphere, the dissolved gas content is will increase as the medium cools and time goes by. Studies in the USP laboratory have demonstrated that good performance verification testing (PVT) results will not likely be obtained if the dissolved oxygen (as a marker for dissolved atmospheric gases) is much greater than 6 ppm. It has been our experience that degassing water at a temperature between 41 and 45 °C can produce an acceptable medium for the PVT. In acknowledgment of the fact that the degassed medium must cool to the test temperature before the start of the dissolution PVT experiment, we performed studies in our lab to determine a minimum degassing temperature that would allow the medium to remain above the 37 °C test temperature during subsequent vacuum treatment, transfer to the vessels, and placement in the dissolution test assembly while requiring minimal additional time for thermal equilibration. It was observed that 41 °C fulfills that expectation. However, in acknowledgment of the ability of a laboratory to control the medium temperature during its preparation, a range of 41-45 °C is practicable. Remember that once the medium cools down to 37 °C and time passes, the dissolved gas content will increase to saturated values. Therefore, before the dissolution run, care must be taken not to let the medium stand for any length of time and not to stir or agitate (such as by pouring the medium with no regard to the potential for mixing atmospheric gas) the medium unnecessarily.