dx.doi.org/10.14227/DT100303P22

Question and Answer Section

William Brown and Margareth Marques
The following questions have been submitted by readers of Dissolution Technologies. Margareth Marques, Ph.D. and Will Brown, United States Phamacopeia, authored responses to each of the questions.
*Note: These are opinions and interpretations of the authors, and are not necessarily the official viewpoints of the USP

Email for correspondence: web@usp.org

Q ­ I would like to have an explanation with details on how to perform manual sampling during the dissolution test.
A ­ Sampling should be done with the paddles or baskets in motion. The sample is withdrawn from a zone midway between the surface of the dissolution medium and the top of the rotating basket or blade, not less than 1 cm from the vessel wall. Our suggestion is to use large volume glass or plastic syringes with a stainless steel cannula bent in an L shape and available from a number of sources. This facilitates taking the sample at the appropriate sampling point even for vessels at the back of the equipment.You should determine if there is any interference from the syringe material in the analysis. The specimens are to be withdrawn at the stated times,within a tolerance of ±2%. In order to comply with this time requirement,you may introduce the capsules or tablets into the dissolution vessels at staggered times leaving enough time between sample introductions to compensate for the sample withdrawal and filtration. In some cases,you may need to contact the manufacturer of your dissolution equipment to determine any modifications needed to allow the motion of the baskets or paddles to be started after the sample is introduced into each vessel. The samples should be filtered promptly after being removed. The filter should be evaluated for adsorption of the drug substance and for possible leaching of substances that could interfere with the analysis. Analysts need to be very well trained to ensure the reproducibility of the sampling procedure.

Q ­ I deaerated the dissolution medium according to the procedure described in the document that comes with the USP calibrator tablets.How should I transfer this hot medium to the dissolution vessel? Can I use graduated cylinders or volumetric flasks?
A ­ Since the volume of medium specified in the USP dissolution test is measured at room temperature, and the test is performed with medium that has been heated,you must take into account the density of the medium at room temperature. One way to determine the rough density is to weigh the amount of medium that will fill a 1- or 2-Liter volumetric flask to the mark. The approximate density is the ratio of the mass of room temperature medium to the nominal volume of the flask. Multiply the target volume (generally 900 mL) by the ratio expressed as grams per mL to get the mass of hot medium needed.

Q ­ What procedure is used to validate filters used in the dissolution test?
A ­ Either a 100% standard solution or a dissolved sample solution (e.g.,prepared as a typical dissolution sample in a dissolution vessel or as a sample in a beaker that is stirred with a magnetic stirrer for one hour) is filtered at least three times. For the standard solution,compare the filtered solution response to that of the unfiltered solution. For the dissolved sample,filter a portion of the dissolved sample solution and compare response to that of a portion of the sample solution which has been centrifuged. For the filter to be acceptable for use,the response of the filtered portions should be within 98% to 102% of the response of the unfiltered standard solution (first case) or the centrifuged sample solution (second case). If the filters are unacceptable,the evaluation might be repeated,with a volume of filtrate discarded before collection of the sample for analysis begins. Based on the acceptability of the filtered sample,the discard volume may need to be increased. If an acceptable discard volume is not found,then a change to another type of filter is recommended.

Q ­ Is the replacement of dissolution medium after each sampling mandatory? How should I do this replacement? How can not replacing the dissolution medium influence the dissolution profile? Is there any calculation I should use to make any corrections?
A ­ USP recommends that the sample volume taken be replaced with fresh medium heated to 37°±0.5° for multiple sampling times. If this is done,the calculation of amount dissolved will use the nominal volume and will account for the mass of active taken in previous aliquots. The replacement volume should be added carefully so as not to disturb the flow in the vessel or the dosage form residue.

USP also allows the test to be continued without replacement of the sampled volume if replacement can be shown to be unnecessary. In this case the calculation will reflect the amount of medium actually present at sampling. The mass of active taken in previous aliquots must still be included in the calculation of cumulative amount released.

The dissolution rate of drug from its dosage form can be influenced by the energy available from the moving medium. The energy input from the apparatus should remain constant. However,with less mass of medium and different boundary conditions (lower medium surface) the energy available will change.The mixing pattern of the solution can also change. The effect of these changes must be evaluated.