In Situ
Dissolution Testing Using
a UV Fiber Optic Probe Dissolution System
Kevin C. Bynum*, Erik
Kraft, John Pocreva, Emil W. Ciurczak, Philip Palermo
Purdue Pharm, L.P., International R&D, Ardsley, NY
Purdue Pharm, L.P., International R&D, Ardsley, NY
* e-mail address for correspondence: kevin.bynum@pharma.com
Introduction
Dissolution testing is a time-consuming
and labor-intensive test procedure. The technique requires a sample
of dissolution medium be removed from each of the test vessels
at a specified time point(s). This sampling procedure can be handled
either manually or by means of an automated sipper mechanism.
Each method has its own inherent errors associated with it. The
manual technique is subject to analyst variability. Two of the
most common errors associated with the manual sampling technique
are improper sampling within the vessel and timing errors in sampling.
Additionally, this is the most labor-intensive technique and is
especially taxing for controlled-release formulations, which often
require 24-hour testing periods. Automatic sampling techniques
can save analyst labor, but still require a good deal of maintenance
and set-up time. Also, automated sipping mechanisms are complex
and contain many moving parts that can malfunction, resulting
in sampling errors. Sipper mechanisms are prone to sample contamination,
carryover, bubbles, dilution, filter clogging, and tubing leaks.
We have developed
a simple solution to these problems. We have decided to eliminate
this sipping step and conduct an in-situ spectroscopic measurement
in the dissolution vessels. We pipe the light (not the liquid),
to a series of 12 dip-type transflectance probes (C-Technologies,
Cedar Knolls, NJ) that reside in the 12 vessels throughout the
dissolution. A pair of deuterium lamps is used as light sources.
After interacting with sample, the light is piped to a series
of 12 Photo-Diode Array (PDA) spectrophotometers. The spectrometers
are Zeiss MMS UV (190-385nm) modules obtained from Hellma (Forest
Hills, NY). At the start of each run, the system Figure
1 collects
"dark" spectra, 100% transmittance spectra, and standard
scans. It also runs a fully automated system suitability procedure.
Once this procedure passes the specified limits, the probes are
placed in the vessels Figure 2, the dosage units are dropped
into the vessels, and the system acquires dissolution data without
further analyst intervention. A full absorbance scan is collected
in real time from each vessel at time intervals ranging from minutes
(for a 24-hour dissolution) to seconds (for immediate-release
dissolution). The system is controlled by an in-house developed
and validated software package. Once the full spectrum is collected
and stored on our computer network, the resultant "percent
dissolved" is calculated in real-time and displayed for the
user on the monitor. When the run is complete, the final report
is prepared by simply pushing a button; no post run calculations
are required.
To demonstrate the performance of the system, three dissolution
experiments were conducted. The first consisted of a number of
immediate release formulations, where the amount of excipent was
adjusted to control the immediate release characteristics. The
second experiment was a twelve-hour dissolution, conducted on
three samples of a controlled release formulation. The third experiment
was a 24-hour dissolution conducted on twelve samples of a once-a-day
controled release formulation.
Immediate
Release Experiment
Four separate samples were used to study the ability of the system
to profile immediate release formulations. The system was set
up to acquire a full UV spectrum from each of the four vessels
every 10 seconds. A Hanson SR-8 dissolution apparatus was fitted
with paddles (USP Apparatus II) rotating at 100 RPM and was equilibrated
at 37° C for one hour. The fiber optic probe system was set
up by scanning dark, 100% transmittance, and standard scans on
all four probes. The probes were then placed into the vessels,
and the samples inserted. The system acquired an absorbance scan
every 10 seconds for the duration of the experiment. A single-point
baseline correction was utilized to eliminate small baseline offsets.
The resultant "percent dissolved" was calculated by
the software in real time and the dissolution curve displayed
to the analyst as the experiment proceeded. At the end of the
experiment, the results were immediately ready for the analyst
to print and review.
Controlled
Release Experiment-12 hour release
Three samples of controlled-release analgesic formulations were
used to demonstrate the ability of the system to monitor long-term
dissolution experiments. The experiment was run on a Hanson SR-8
dissolution apparatus configured with baskets (USP Apparatus I)
operating at 100 RPM. The system was equilibrated at 37° C
for one hour. The fiber-optic probe system was set up by scanning
dark, 100% transmittance, and standard scans on the 3 probes.
The probes were placed into the vessels and the samples added.
The system acquired an absorbance scan from all of the vessels
at ten-minute intervals for 12 hours. A novel (proprietary) algorithm
was used to correct for the particulate matter that was present
from the controlled release matrix. The percent dissolved curve
was calculated in real time and was compared with data obtained
from the HPLC method that the fiber optic technique is replacing.
Controlled
Release Experiment-24 hour release
Twelve samples of a once-a-day controlled-release analgesic formulation
were used to demonstrate the utility of the system for routine
stability testing. This experiment was conducted on two, six vessel
Hanson SR-6 dissolution apparatus configured with paddles operating
at 100 RPM. The system was equilibrated at 37° C for one hour.
The fiber-optic probe system was set up by scanning dark, 100%
transmittance, and standard scans on all 12 probes. The probes
were placed into the vessels and the samples added. The system
acquired an absorbance scan from all of the vessels at ten-minute
intervals for 24 hours. A single-point baseline correction was
utilized to eliminate small baseline offsets. The percent dissolved
curve was calculated in real time and was compared with data obtained
from the HPLC method that the fiber optic technique is replacing.
Results
The results of the immediate release experiments are presented
in Figure 3.
It can be seen that the data-rich graphs show both the gross and
subtle differences between formulation dissolution profiles.
The 12-hour controlled-release experiment in Figure 4 shows the versatility of
the method in that specifications for percent dissolved at nearly
any time point may now be set. In most dissolution trials, fewer
time points are sampled, making "off-time" specifications
difficult. That is, if the samples were drawn at four, eight,
and 12 hours, a percent dissolved figure for 10 hours would have
to be extrapolated in current systems. With the fiber optic system,
the software would simply print out the 10-hour result.
In addition, the close agreement of the in-situ probe data with
the HPLC data indicates that the resident probes do not effect
the release of the active component from this formulation.
The 24-hour controlled-release
experiment in Figure
5 shows that
the system works well under routine stability conditions. The
close correlation of the data with the HPLC results demonstrates
the accuracy of this technique.
Discussion
As Figure 3
illustrates, the in-situ method also allows for rapid data acquisition.
Our system has been able to acquire scans on all 12 channels once
every 12 seconds. This rapid-scanning procedure has allowed us
to profile the release characteristics of immediate and explosive
release formulations.
As is shown in Figure
4, the
UV Fiber Optic Probe Dissolution System generates a continuous
dissolution curve consistent with the external assay performed
by HPLC. This accuracy is also demonstrated in Figure 5, which demonstrates the
application of the system for the analysis of 6 or 12 sample lots
over a 24 hour period.
The increase in "data density" during the dissolution
experiment increases the statistical significance of the data
collected. This can be very useful when using this data to demonstrate
an In-Vivo In-Vitro Correlation. The increase in data allows for
predictive modeling of dissolution curves.
The system is much
easier to set up than systems with sippers. Once the system passes
the system suitability portion of the experiment, the system will
run continuously for the specified dissolution time. This ruggedness
is due to the solid state nature of the system. Since there are
no moving parts, breakdowns are rare.
The absence of a
chromatographic step makes the setup quick and easy while eliminating
any error associated with HPLC. It takes the average analyst about
two hours to set up a dissolution experiment (including all reagent
preparation). The system also performs all calculations in real
time; which eliminates the need for post-dissolution paperwork,
while allowing for post-run review of the data.
Conclusion
A UV Fiber Optic Probe Dissolution System has been developed for
the analysis of solid dosage forms. The system uses 12 dip-type
fiber-optic probes coupled to 12 separate PDA spectrophotometers
to acquire continuous dissolution curves in real time. The system
is applicable to the analysis of both immediate and controlled-release
formulations. The system is accurate, quicker, and easier to set
up when compared with conventional HPLC or UV-sipper systems.
The data is acquired, calculated, and secured using an in-house
developed and validated software package. The software package
uses a novel scatter-correction algorithm to generate accurate
dissolution curves in a turbid medium. The data that the system
generates is as accurate as the HPLC methods that it is replacing.