William Brown and Margareth Marques
The following questions
have been submitted by readers of Dissolution Technologies. Margareth
Marques, Ph.D. and Will Brown, United States Phamacopeia, authored
responses to each of the questions.
*Note: These are opinions and interpretations of the authors,
and are not necessarily the official viewpoints of the USP
Email for correspondence:
web@usp.org
Q
Is there a table that compares/contrasts the
various USP dissolution instruments?
A
See Table 1.
Q
Is there any harmonization on dissolution testing
equipment for use in the drug-elution device field?
A
Harmonization of the dissolution chapters in the European
Pharmacopoeia,Japanese Pharmacopoeia,and US
Pharmacopeia has produced agreement on the descriptions
for Apparatuses 1�4 (basket,paddle, reciprocating
cylinder, and flow-through cell,respectively). If drug release
from novel drug-eluting devices can be evaluated using
any of these four apparatuses, then we would be able to
answer,yes. The USP Biopharmaceutics Expert Committee
has initiated work by several newly formed ad hoc Advisory
Panels to evaluate performance testing of novel dosage
forms. The Panels are charged with recommending new
apparatuses if the testing cannot be performed using those
already described in USP. If new apparatuses are included in
USP,there is a possibility for harmonization.
Q
How do you calibrate small volume vessels?
A
By calibration you mean demonstrate suitability for use.
Such a demonstration will require a full understanding of the
critical operating and dimensional variables at work in testing
with small-volume vessels.Once an understanding is attained
of the effect that deviations in these variables produce on
dissolution results,we can start to find ways to determine that
the system is suitable.
Q
How to decide if the medium should be deaerated
or not?
A
Dissolution profiles should be obtained for the same
sample using deaerated and non-deaerated media.The
profiles should be compared and the condition that gives
better discrimination should be considered.
Q
If a particular type of dissolution equipment is
used with both manual and automatic sampling
how do we perform its calibration?
A
Automated sampling brings variables into play that,
strictly, are not controlled by calibration. Calibration is
intended to control the variables associated with the
production of drug solution in the medium by the apparatus.
In the dissolution procedure, automated sampling
is downstream of that process. In our opinion, control of
the timing of sampling, carryover in any sampling lines,
pump flow rate to the spectrophotometer, volume
removed, and other variables in the automated sampling
system can be better performed by other means.We
recommend that calibration be done with manual
sampling.
Q
Should the placement of baskets and paddles be
fixed in a given equipment?
A
As far as we know there are no guidances or guidelines
stating that the paddle or basket or shaft should have
dedicated positions in any dissolution equipment.We
know that there are several companies using this
approach to facilitate the evaluation of out of specifications
results or any other deviation.
Q
We are evaluating a dissolution test for orally
disintegrating tablets and we are getting very variable
results because the tablets float.We cannot use
sinkers because the tablets are very fragile.Do you
have any suggestions to solve or minimize this
problem?
A
There are some options you can try:1 - the use of USP
apparatus 1 (basket);2 � the use alternative sinker given in the
harmonized Dissolution General Chapter (see page 205 of the
Pharmacopeial Forum 31(1) [Jan-Feb 2005]) or see USP
monograph for Nifedipine Extended-release Tablets on page
1377 of USP 28),this sinker is not tightly wrapped around the
dosage form;3 � you can try a basket similar to the one
described in USP monograph for Felodipine Extendedrelease
Tablets (pages 808 � 810 of USP 28). In any case,you
will need to verify the suitability of any of these devices in
practice.
Q
What should be the next steps when we have f1
and f2 not passing the acceptance criteria for two
products that are being compared?
A
Dissolution profile comparison is very helpful when evaluating
post-approval changes and biowaivers. It helps to
assure similarity in product performance and signals
bioinequivalence.To assure similarity in product performance,
it is important to know how similar the two curves are.
Thus,the f2 comparison is very useful (See Dissolution profile
comparison using similarity factor,f2. Dissol.Techn. 6(3):15,
1999). If the comparison shows that the dissolution profiles
are not similar,two approaches are possible:1 � make small
changes in the formulation and or manufacturing process
and compare the profiles of the product after the changes or,
2 - carry out a bioequivalence study with a small population
so see if the in vivo performance shows that the products are
really different.
Q
If I have one vessel out of requirement using the
USP Calibrator Tablets,can I continue the testing
through S2 and S3?
A
The Acceptance Table in the USP general chapter <711>
Dissolution is not applicable to the calibration of dissolution
equipment.The procedure and acceptance criteria for the
calibration are available at http://www.usp.org/standards/calibrators.html. When you have one vessel giving results out
of the ranges for calibrator tablets,an investigation should be
carried out to find the possible causes of this deviation.These
causes must be corrected and a new calibration,including all
vessels,should be run.
Q
We are developing a dissolution test for a product
with very low strength.The amounts released into
the medium are very small and the quantitation step
is by HPLC.Can we introduce two dosage forms in
each dissolution vessel to increase the amount of
drug released into the medium?
A
Introducing more than one dosage form in each dissolution
vessel is not a good approach.To solve this problem
related to very low amounts of drug dissolved in the medium
you have some options:1 - decrease the volume of dissolution
medium.Although 200 mL vessels are not compendial
they can be used together with mini paddles.They should be
properly calibrated (see question 3 above);2 � you can
change the chromatographic system to increase the sensitivity
of the method;3 � you can use another analytical technique
with better sensitivity;4 � you can increase the amount
of sample injected into the HPLC column,making the appropriate
modifications in your chromatographic system.
Q
Can we modify the quantitation step of an USP
dissolution test but keeping all dissolution conditions
(apparatus,speed,medium,etc) the same?
A
Yes,you can.The USP General Notices (page 7 of USP 28)
states that �Compliance may be determined also by the use of
alternative methods,chosen for advantages in accuracy,
sensitivity,precision,selectivity,or adaptability to automation
or computerized data reduction or in other special circumstances.
Such alternative or automated procedures or
methods shall be validated.�
Q
We are developing a dissolution test for a
powder material and we are having problems
with the sampling.Powder is being adsorbed to
the filter of the sampling probe.Do you have any
suggestions?
A
Automatic sampling is not suitable for all dissolution tests.
In some cases,results are more precise and reproducible
using manual sampling because you can use pre-filtration
using a coarser material,and subsequently filter through a
material with a smaller porosity.You should also evaluate
different filter materials to find one where your powder might
not be adsorbed.Another possible option is to use the USP
Apparatus 4 (flow-through cell).
Editors Note: The following Question deals specifically with USP Apparatus 3.We consulted with a technical adviser of a manufacturer of this equipment,Bryan Crist,who supplied the response.His email address is bryan.crist@Varianinc.com
Q
My question is specific to the type 3 dissolution
apparatus.The rim of the outer medium tube
(250�300mL dissolution vessel) is extremely susceptible
to wear and tear.All our vessels are chipped or
cracked at the rims.Is there any product on the
market such as a rubber covering for the rim that
could prolong the life of these vessels? This would also
solve the problem of the sampler tube getting caught
in the chips and cracks of the rim during a row change.
A
Rims may become chipped when struck by the evaporation
covers and cannulas if the vessels are allowed to float in
the media.Vessels should only be in the apparatus when they
contain dissolution medium,which is necessary to keep them
at the proper level. Laboratories that are consistent with this
practice have very few cracked vessel rims. Also,if a low
volume of dissolution medium is used in the vessel,the circulator
water must be at a level low enough to prevent the
vessels from becoming buoyant and struck by the cannulas as
they transverse from row to row.Thank you for your suggestion
regarding a covered rim to prolong the life of the vessels.