dx.doi.org/10.14227/DT160409P42

Question and Answer Section - November 2009

William Brown and Margareth Marques
The following questions have been submitted by readers of Dissolution Technologies. Margareth Marques, Ph. D. and Will Brown, United States Phamacopeia, authored responses to each of the questions.
*Note: These are opinions and interpretations of the authors, and are not necessarily the official viewpoints of the USP

Email for correspondence: web@usp.org

Q What are the required number of samples when evaluating dissolution of multi-dose and single-dose units of oral suspensions?
A The required number of individual dissolution results is defined by the acceptance table. The current USP general chapter <711> Dissolution does not specify the number of containers of suspension, or powder or granulate for suspension, that should be used as the source of the samples for dissolution testing. The recommendation is to transfer to the vessel an amount of product equivalent to one dose or the amount specified in the particular monograph. The product must be reconstituted as directed on the label instructions. If the product has several different doses (e.g., dosing according to age or body weight), the test should be developed for the highest dose.

Q On what basis is the total volume of dissolution medium for a particular product defined?
A The volume of medium is usually determined so that sink conditions are maintained. The solubility of the drug substance in different media should be determined. Based on this information, the volume of medium needed to obtain a saturated solution, considering the highest dose that will be marketed, should be defined. The recommendation is to work with volumes at least three times the one necessary to obtain a saturated solution. In some cases, it may be appropriate to work with higher volumes, such as 5 or even 10 times the saturation volume based on other considerations. The definition of the medium volume is determined case by case and should be part of a discriminative procedure for a particular product. Keep in mind that there are some instances where the most discriminative condition may not conform to sink conditions.

Q How is the apparatus rotation speed selected for a particular drug product?
A Dissolution profiles should be obtained at different rotation speeds, and the most discriminative condition should be selected. Ideally, the dissolution results will reflect in vivo performance. In most cases, the recommended rotation speed is between 50 and 100 rpm. Some products, like suspensions, may use rotation speeds as low as 25 rpm. In general, rotation speeds less than 25 rpm or greater than 150 rpm are not recommended. The selection of the rotation speed is done case by case.

Q We are relocating several dissolution apparatus. Some are going to be moved to a new position on the same bench, others are going to be moved to other benches in the same lab. Should we go through the performance verification for all of these test assemblies?
A Every time a dissolution apparatus is moved or serviced, a complete requalification of the equipment is recommended. During the relocation of the equipment, parts of it may be dislodged, unleveled, or even damaged. Sources of vibration at the new location are another critical point. If servicing is needed, parts of the equipment may be replaced. In any case, it is necessary to verify that the performance of the equipment is still within acceptable limits. Recommendations on how to qualify a dissolution apparatus are available at http://www.usp.org/pdf/EN/dissolutionProcedureToolkit2007-10-04.pdf

Q Are 2-L dissolution vessels widely accepted by regulatory agencies around the world?
A The USP general chapter <711> Dissolution was harmonized for the U.S. Pharmacopeia (USP), the European Pharmacopoeia, and the Japanese Pharmacopoeia, but there are certain parts of the text, specially marked, that were not harmonized. One example of this situation is the volume of dissolution vessels. Only 1-L vessels are harmonized among the three pharmacopeias. It does not necessarily mean that 2-L and 4-L will not be accepted, but good justification for using high volumes of dissolution media should be part of the documentation submitted to the regulatory agency in the region where the product will be marketed. The most common use for a high volume of dissolution medium is for poorly soluble drugs. There are some USP monographs and products approved by FDA that use volumes higher than 1 L (see http://www.accessdata.fda.gov/scripts/cder/dissolution ).

Q If both manual and automated procedures are used for sampling in dissolution testing, both of them should be evaluated in the performance verification of the equipment. How is �automated system" defined? Is an online HPLC UV detector part of the system, or is it only the dissolution sampling station?
A A manual system is one in which the dosage units are introduced by hand, the timing is by an external chronometer, the test solution is taken up by hand using a syringe or other similar device, and manipulations required for spectrophotometric or chromatographic analysis are done by hand. Some degree of automation is inferred for systems where any of the above processes are under immediate control of a microprocessor. The Performance Verification Test (PVT) is a measure of the suitability of a dissolution test assembly to produce acceptably similar dissolution samples in comparison with laboratories worldwide. All laboratories who participate in the collaborative study conducted by USP to determine the acceptance ranges for each lot of USP Reference Standard tablets are expected to perform the testing manually. Automated sampling brings additional variables in sample preparation. Although manual and automated sampling should produce similar results, USP does not specifically certify that its ranges are suitable for automati- cally sampled systems.

Q According to the General Notices, numbers are not rounded until the final calculation for the reportable values. The acceptance tables in <711> Dissolution have criteria for several levels such as S1, S2 or L1, L2, L3. Dissolution values are obtained from each unit tested. When calculating the aver- age for L2 or L3, is the average calculated from the rounded individual number results reported or from the unrounded results?
A Use section 7.20 of the General Notices as a guide on rounding. For the limit expressions in <711> Dissolution, it is apparent that several limits may be encountered when working through the stages (or levels) given in Acceptance Tables 1, 2, 3, or 4. At each stage of all of the tables, individual unit values need to be compared with limits that are given at a whole-number level of signifi- cance. Therefore, the values are rounded to whole num- bers for that comparison. Where an average of unit values is needed, the original unrounded data are used to compute the average and then rounded to the whole- number level of significance of the limit.