Mechanical
Versus Chemical Dissolution Calibration
Tahseen Mirza, Ph.D.
United States Pharmacopeial
Convention, Rockville, MD
* e-mail address for correspondence: tm@usp.org
This paper is intended to serve as a response
to the article titled, "Recent Innovation in Dissolution
Calibration," written by Mary Oates, Ph.D., in the August
issue of Dissolution Technologies. The aforementioned article
summarized the recent PhRMA Dissolution Subcommittee's collaborative
studies/activities aimed at total elimination of the use of calibrator
tablets in dissolution calibration. The article recommends mechanical
calibration as a potential replacement for calibrator tablets.
Since such a radical deviation has faced skepticism, PhRMA on
an interim basis has recommended the elimination of one calibrator
tablet (salicylic acid) and replace it with a more thorough mechanical
calibration.
It is important to bear in mind
that the USP Committee of Revision, and more particularly, the
USP Subcommittee on Dissolution and Bioavailability (DBA) is the
body that has the responsibility for making changes to the calibration
procedures for Dissolution Apparatus by making appropriate revisions
to the USP General Chapter <711>. PhRMA or any other interested
party can make recommendations to the DBA for their consideration
as a possible revision. The USP General Chapter <711> contains
the dissolution test that is required to determine compliance
with the dissolution requirements in an individual monograph.
These requirements are enforceable by regulatory agencies in countries
where USP is recognized as the official source of public standards.
The DBA Subcommittee in their last meeting reviewed PhRMA's recommendations.
Members of the DBA Subcommittee concluded that the data submitted
by PhRMA was not persuasive and have unanimously voted against
it (1). Therefore, the calibration procedures for USP Apparatuses
1 and 2 have not been affected and will continue to require the
use of both calibrator tablets; non-disintegrating salicylic acid
and disintegrating prednisone tablets. Mechanical calibration
and visual inspection of apparatus must precede chemical calibration
using the USP calibrator tablets. It is the opinion of the DBA
that mechanical calibration alone is not sufficient to ensure
that the system is suitable for use prior to compendial dissolution
testing.
Mechanical calibration, as defined
in Dr. Oates' article is "the measurement of physical parameters
of the dissolution bath which, when stringently controlled, provide
evidence of the proper operation of the apparatus." Therefore,
by definition mechanical calibration is a test of the apparatus
only. It provides no assurance that the other important components
of the system: the analyst, the analytical instrument and analytical
procedure are performing satisfactorily. Nor does it assure that
the interactive effect of the apparatus variables have a detrimental
effect on the dissolution rates.
The USP 24/NF 19 contains 586 dissolution
tests and 31 drug release tests (dissolution tests for extended-release
dosage forms). The Biopharmaceutical Classification System and
the various FDA guidances for biowaivers on the basis of dissolution
results are some of the examples of the continued dependence of
the regulatory agencies on the test. It is therefore fair to assume
that the dissolution test is not only here to stay, but is expected
to find more uses and applications in the future. In this regard,
USP has an obligation and responsibility to ensure that all the
monograph dissolution tests are performed properly. System suitability
testing using calibrator tablets is the most effective means by
which USP attempts to provide such assurance.
USP is followed in approximately
40 countries. The level of analyst training and the availability
of good quality dissolution equipment varies greatly in these
countries. A survey conducted by USP of 17,886 tablets submitted
by 79 companies revealed that there is a distinctly higher failure
rate outside the United States (2). In another reported collaborative
study, 13 out of 23 participating laboratories failed to meet
the USP System Suitability test (3). In order for PhRMA's recommendations
to be adopted by the USP, and applied as a global standard, representative
international laboratories must be included in the collaborative
studies.
Calibration of dissolution apparatus
is a system suitability test. It is comprised of three critical
components; the analyst, the dissolution apparatus and the analytical
procedure/instrument. In order for the test to be performed properly,
these three components have to interact optimally together; otherwise
the dissolution results can be misleading. Each of the components
is affected by its own set of variables. The apparatus variables
include belt tightness, current surges, coplanarity of the base
plate and top plate, shaft verticality, shaft wobble, basket imperfections,
basket cleanliness, vessel cleanliness, vessel imperfections,
vibration, and temperature fluctuations. Some of these variables
such as vibration, vessel cleanliness, basket cleanliness and
vessel imperfections are extremely difficult to measure and control.
Inability to properly measure vibration levels at various points
within an apparatus is the main reason why calibrator tablets
were originally developed. Even to this day there is no convenient
and inexpensive means by which vibration can be properly quantitated
at the point of impact such as the bottom of the vessel where
the tablet originally settles prior to disintegration and dissolution.
The analyst training and technique are also very critical to the
test. Proper apparatus set-up, medium degassing, sample introduction,
sampling from the proper zone at the exact time from each vessel,
sample preparation and filtration, standard preparation and handling
can significantly impact dissolution results. A system suitability
test as required in the USP General Chapter <711> provides
assurance that the combination of all these variables does not
bias the results significantly and hence lends credibility to
the compendial dissolution test results. On the contrary, mechanical
calibration focuses narrowly on the apparatus variables alone
some of which can not be measured accurately.
It will be an almost impossible task to develop one calibrator tablet that would mimic all the ever-increasing variety of dosage forms. Prednisone tablets and salicylic acid tablets were developed to represent the disintegrating and non-disintegrating dosage forms, respectively. Despite the availability of a large number of extended-release dosage forms in the marketplace, we still do not have a calibrator tablet that mimics the dissolution of extended-release dosage forms in Apparatus 1 and 2. Elimination of calibrator tablets in this constantly evolving area of drug-delivery will not be a wise choice. It is the primary goal of the USP to safeguard public health by setting good public standards. It is believed that the elimination or further reduction in the usage of calibrator tablets in dissolution calibration is not in the best interest of public health.
1 Dr. Tahseen Mirza is the Liaison
to the USP Subcommittee on Dissolution and Bioavailability.
Comments should be sent to:
e-mail: tm@usp.org, Phone: (301)816-8201
Fax: (301)816-8373
2 Opinions expressed in this article are that of the author Dr.
Tahseen Mirza and not necessarily those of the USP.
References
1. T. Mirza and T. Foster, Proceedings of the Pharmacopeial Forum
25 (6); 1999.
2. B.B. Hubert, V. A. Gray and T.L. Cecil, USP Survey of Calibrator
Results æ Lot K Sa licylic Acid Tablets and Lot J Prednisone
Tablets, Pharmacopeial Forum 20 (6), 8574, 1994.
3. S.A. Qureshi and I.J. McGilveray, Typical Variabilty in Drug
Dissolution Testing: Study with USP and FDA Calibrator Tablets
and a Market Drug (Glibenclamide) Product, Euro J Pharm Sci, 7
(1999) 249-258.