William Brown and Margareth Marques
The following questions
have been submitted by readers of Dissolution Technologies. Margareth
Marques, Ph. D. and Will Brown, United States Phamacopeia, authored
responses to each of the questions.
*Note: These are opinions and interpretations of the authors,
and are not necessarily the official viewpoints of the USP
Email for correspondence:
web@usp.org
Q
We have a product that is a suspension with two
active substances, one of which has high solubility
in water. For the manufacturing of this product, a
solution of the highly soluble drug is prepared first,
and then the poorly soluble active is added followed
by all the other excipients, resulting in a suspension.
Is it necessary to develop a dissolution test for both
active substances or only for the poorly soluble one?
A
You will need to verify experimentally how the highly
soluble drug substance is affected by the formulation by do-ing a dissolution profile along with the other drug substance.
Depending on the behavior of the two active substances
in the dissolution test, you will decide if you are going to
control the dissolution of both of them or only the poorly
soluble one.
Q
We are running the dissolution test for an ex-tended-release tablet, and the acceptance criteria
at the first time point is not more than 25% of the
drug substance dissolved. How should the level L2
from Acceptance Table 2 in the USP General Chapter
<711> Dissolution be applied for this time point?
A
To apply the levels L2 and L3 of the Acceptance Table 2,
you need a range. In the case of the example mentioned, the
range will be 0—25%. For the level L2, the new range will be
from (0% of the label claim minus 10% of the label claim) to
(25% of the label claim plus 10% of the label claim) resulting
in a new range of -10—35%. As dissolution results cannot be
negative, the new acceptance criterion will be from 0% to
35% of the label claim. The same logic should be applied to
the other time points and to level L3.
Q
Are there any guidelines on the use of peak vessels?
A
Peak vessels are not described in the dissolution general
chapter in part because their geometry has not been stan-dardized. They
can only be used with appropriate justifica-tion. One commonly used
justification is the elimination of
variability due to the formation of a cone of sediment at the
bottom of the vessel during paddle dissolution testing. In
most cases, the problem of formation of a cone of material
at the bottom of the vessel can be eliminated or reduced by
increasing the speed of the paddle. If peak vessels are going
to be used, it is necessary to demonstrate that they bring an
advantage to the method that cannot be achieved with the
standard apparatus. In the USP, there are only two mono-graphs where
peak vessels are used, Galantamine Tablets,
Dissolution Test 3, and Praziquantel Tablets, veterinary use.
Q
What are the differences and advantages of USP
Apparatus 3 and USP Apparatus 4?
A
Apparatus 3 provides a high turbulence that may be
useful with dosage forms such as liquid-filled capsules with
lipophilic filling. It has a limitation that not more than 250 mL
of dissolution medium can be used in each tube, but allows
testing over six tubes. The medium can be modified in each
tube. If testing in multiple tubes is done, the dosage form
should be contained within the mesh at the bottom. For this
reason, granules, pellets, and matrix or layered tablets are
typical samples.
Apparatus 4 allows a broad range of volumes. In the closed-loop configuration, a minimal amount of medium can be used. The open-loop configuration allows the use of an unlimited volume of fresh medium. With modern pumps, flows from 1 to 2 mL/min to as much as 50 mL/min can be achieved. By switching reservoirs, the medium composition can be changed. The large volume available in the open-loop configuration makes this apparatus useful in testing extend-ed-release implants or products with poorly soluble drugs.
Q
The regulatory agency asked our company to
develop a dissolution test for a product that contains
an active substance that is nearly insoluble in water.
It has local action on the GI and negligible systemic
availability. Why should a dissolution method be
developed? How should a dissolution method be
developed for this kind of product?
A
The drug will likely have to be in solution to provide
pharmacological action even though it is not absorbed
into the systemic circulation. That being the case, it should
have some solubility within the physiologic pH range,
roughly pH 1—7.6. Start by evaluating the solubility over
that range. From that data, you might discover a possible dissolution
medium. The use of basket or paddle
apparatus and 900—1000 mL are typical and would be a
good starting point. If dissolution at a reasonable level
cannot be achieved, then it is possible that the addition
of a surfactant to the medium may be needed. Some
drugs rely on naturally occurring surfactants to dissolve
in the GI tract. At the end of the process with dissolution
data in hand, you will still need to justify the conditions
chosen. Some level of physiological relevance as well as
discriminatory power are important aspects of a dissolu-tion method.
Q
One compendial monograph has a dissolution
test with a tolerance of NLT 75% (Q) of the label claim
dissolved in a certain time. However, in USP General
Chapter <711> Dissolution, the acceptance criterion
is Q + 5%. Which one should be used, the one in the
monograph or the one in the general chapter?
A
In dissolution tests found in USPmonographs, the
tolerances are typically presented in the format ìnot less
than X% (Q)î. The Qvalue is a variable that is substituted
into the acceptance criterion given in Acceptance Table
1. Acceptance Table 1 allows the test to stop after the first
stage (six units tested) if all results are not less than Qplus
5%. In your example, that would be 80%. However, for
the later stages of the test, the average of the units tested
must not be less than Q.
Q
We are developing a new product that contains
an active substance in the hydrochloride form that
is not soluble and forms a precipitate in basic pH.
We are conducting some evaluations using fasted-state simulated intestinal fluid (pH 6.5) and fed-state
simulated intestinal fluid (pH 5.5), and the drug
substance precipitates immediately in both media.
Do you have any suggestions?
A
The use of dissolution media that simulate biological
fluids is a useful tool to gather more information about how
the dosage form could perform in vivo. The use of these
types of dissolution media is not mandatory and does not
have application in routine quality control. The information
obtained with dissolution testing using simulated biological
fluids should be shared with the formulation development
team to help them seek formulation alternatives to prevent
or minimize the precipitation that may occur in vivo.
Q
Can Acceptance Table 3 from the USP General
Chapter <711> Dissolution be used for the in-pro-cess dissolution testing of delayed-release pellets?
They are filled into capsules at the clientís facility.
A
The USPGeneral Chapter <711> Dissolution is silent on
the dissolution testing for in-process controls. The acceptance
tables are not product specific. Acceptance Table 3
gives general criteria for acid resistance testing. The instructions
stated in the chapter can be adapted for this particular
application. Typically, an amount of pellets equivalent to the
target capsule contents will be tested.
Q
In some disintegration procedures for delayed-release capsules,
the use of disks is recommended. We
are developing a new disintegration method for this
type of dosage form, and we would like to know if we
should use disks or if we can use sinkers to prevent
the capsules from getting out of the tubes.
A
The use of sinkers is not part of disintegration testing. If
necessary and with justification, the disks described in the
USPGeneral Chapter <701> Disintegration can be used.
The description of the setup of the disintegration test, if
followed, will prevent the immersion of the tops of the
tubes. Therefore, floating capsules should not be able to
escape.
Q
What are the effects on the results when the dos-age form does not
drop to the center of the vessel in the paddle apparatus?
A
There is no general answer to this type of question. The
hydrodynamics within the paddle apparatus has been
described in the literature. Shear is at a minimal level in the
region below the tip of the paddle shaft. Experimental results
have shown that tablets falling to rest away from the vessel
center may have higher dissolution results than tablets at the
vessel center. Every attempt should be made to initiate the
test in a standard fashion. Therefore, the dosage form should
fall to rest at the center of the vessel bottom. Sinkers have
been used to facilitate positioning the dosage form at the
bottom of the vessel.
Q
According to the certificate that accompanies USP
Prednisone Tablets RS, if the equipment has fewer
than 12 positions, step 1 should be repeated with
an additional set of tablets. If dissolution apparatus with six and
eight vessels are being evaluated,
should the test be carried out with two different sets
of tablets?
A
The Performance Verification Test (PVT) using USP
Prednisone Tablets RS, lot P1, for test assemblies with not
more than eight vessel positions can be performed either
as a two-stage or a single-stage procedure. The choice of
using the single- or two-stage procedure must be made in
advance of the test. Both procedures can use as many as 2X
tablets, where X is the number of positions. For a six-position
test assembly, 2X would be 12 tablets tested. The two-stage
procedure allows a stop at only 1X if the criteria are met. For
the six-position assembly, that would be after only six tablets
are tested (one tablet in each vessel). For the two-stage procedure,
if the criteria after stage 1 are not met, six additional
tablets are tested.