dx.doi.org/10.14227/DT070200P6

Manual In Situ Fiber Optic Dissolution Analysis in Quality Control
Schatz C.,(1 ) Ulmschneider M.,(1 ) Altermatt R.,(1 ) Marrer S.,(1 ) and Altorfer H.(2)
(1 ) Pharmaceutical Quality Control and Quality Assurance, F. Hoffmann-La Roche Ltd., Basel, Switzerland
(2)
Pharmaceutical Analysis, Swiss Federal Institute of Technology, Zurich, Switzerland


email for correspondence: caspar.schatz@roche.com

Abstract
Manual dissolution testing entails time-consuming sample preparation. Samples withdrawn from the dissolution vessels require filtration, then the clear solutions have to be transferred into cells for quantification by UV/VIS spectroscopy via an absorption reading.
This article describes the validation of a new method that enables the extent of dissolution to be determined without any sample preparation because the active substance concentration is measured directly in the dissolution vessels with a fiber optic probe. Using soft gelatin capsules and tablets as models, the method was validated with regard to linearity, accuracy, precision, specificity and robustness. The analytical results confirmed that the fiber optic quantification method is simple, reliable, reproducible, robust, time-saving, and easy to use in dissolution testing.

 

Introduction
As a compendial requirement the dissolution methods for a given drug product are fixed in terms of medium, temperature, stirring speed, and stirring devices. Any change in a parameter that affects dissolution efficiency, such as paddle speed or dissolution medium, requires extensive experiments to get the approval of the regulatory authorities. Options for reducing the analysis time for manual dissolution tests are therefore very limited. However, the method employed to quantify the amount of active substance dissolved is a key determinant of the analysis time. Usually the quantification is done either by high performance liquid chromatography (HPLC) or by direct UV/VIS spectroscopy. Owing to the additional separation step, HPLC is more time-consuming and more expensive than direct UV/VIS spectroscopy. Since the dissolution testing of almost every solid drug product leads to a turbid solution, sample withdrawal from dissolution vessels must be followed by a filtration step to get rid of undissolved components from the drug product and to eliminate potential interference during quantification. When using direct UV/VIS spectroscopy for quantification, the clear solutions are transferred to cells prior to the absorption reading.

Using spectroscopy instead of chromatography to determine the amount of active substance in solution is one way of reducing the necessary analysis time. Another approach minimizes the time spent on sample preparation prior to manual dissolution testing. This can be achieved by using a fiber optic probe to measure the absorption in the dissolution vessel directly. Since the filtration step is omitted, the analytical method has to compensate for the turbidity effects of undissolved excipients by using two wavelengths. One wavelength - preferably at an absorption maximum - is used for quantification and the other - at a point where none of the excipients shows any absorption - is used for the turbidity correction.

This approach was tested on two different drug products. One was a soft gelatin capsule, containing 200 mg of active substance dissolved in a mono- and diglyceride filling. While performing the dissolution test with these capsules the medium becomes turbid owing to the metal oxide pigments from the capsule shell. The other product examined was a benzodiazepine tablet formulation containing 6 mg of active substance. In this case the turbidity of the dissolution medium is mainly due to microcrystalline cellulose.

Experimental
All dissolution tests were performed with apparatus 2 [1] (Sotax AT 7, Sotax AG, Allschwil, Switzerland). A scanning spectrometer (Varian Cary 50, Varian International AG, Zug, Switzerland) was used with a fiber optic probe (Ultra Mini TS 10 mm + 2 LL UV Li/SMA 974725/18, Hellma GmbH & Co., Müllheim/Baden, Germany) with a 10 mm path length. As shown in Figure 1, the absorption readings with the fiber optic probe were taken via the sampling hole of the dissolution tester about 3 cm below the surface of the medium, with degassing by helium sparging or sonication under vacuum [2].

Figure 1: Experimental arrangement for taking absorption readings using a fiber optic probe.

Soft gelatin capsules
The soft gelatin capsules have a Q-value of 85% after 20 minutes in 900 mL of dissolution medium (aqueous citrate buffer solution, pH 3.0, with 12% Cremophor® EL) at 37.0°C ± 0.5°C stirred at 100 rpm. Except for the robustness determination, all dissolution tests were performed under these conditions.

Method development
Dissolution tests with the drug product, the active substance, and the placebo formulation in dissolution medium were performed in order to establish a suitable wavelength for the turbidity correction.

Methods
In the case of the registered method (validated according to Ref. [3, 4]) an aliquot of about 20 mL was withdrawn from each vessel after 20 minutes. The samples were filtered (pore size 0.45 µm) and the absorption of the clear solutions was measured at room temperature against a dissolution medium blank at 315 nm (A315) with a cell width of 10 mm. The amount of active substance dissolved was calculated as a percentage (%dissolved) using Formula 1, where 4,500 is a factor accounting for the dilution and the conversion to percent and 53.0 is the A (1%, 1 cm) value of the active substance in clear dissolution medium at room temperature.

Formula 1:
%dissolved=A315*4,500/53.0

Using the fiber optic probe, the absorptions at 315 nm (A315) and 380 nm (A380) were measured directly in the vessel. The amount of dissolved active substance in percent (%dissolved) was calculated using Formula 2, where 53.0 is the A (1%, 1 cm) value of the active substance in placebo-spiked dissolution medium at 37.0° ± 0.5°C.

Formula 2:
%dissolved=(A315-A380)*4,500/53.0

Validation of the fiber optic method
The linearity of active substance in placebo-spiked dissolution medium was examined in replicates of six at six concentrations ranging from 5 to 120% dissolved active substance (0.011 ­ 0.267 mg/mL).

A (1%, 1 cm) was calculated with these data, as well as the accuracy and the precision as an internal validation. A method comparison with the registered method for dissolution testing of the soft gelatin capsules was carried out for the additional determination of accuracy and precision. For these experiments six capsules from seven lots were used to determine the amount of active substance dissolved.

The placebo response was determined with six placebo capsules as well as six soft gelatin capsule shells of the drug product. The robustness of the analytical method for determining the amount of active substance dissolved was examined in terms of temperature, paddle speed, and the immersion depth of the fiber optic probe window. In addition, dissolution profiles over 20 minutes were measured with six soft gelatin capsules.

Benzodiazepine tablets
The registered Q-value of the benzodiazepine tablets in 900 mL of simulated gastric fluid without pepsin [1] is 75% after 20 minutes, the vessels being stirred at 50 rpm at 37.0° ± 0.5°C. Accordingly, all dissolution tests were performed with these parameters.

Method development
To establish a turbidity correction wavelength, dissolution tests with the drug product, the active substance, and placebo powder were performed.

Methods
For the registered method (validated according to Ref. [3, 4]) and the fiber optic method, a standard of about 0.0065 mg/mL (cstandard) was prepared and its absorption at 239 nm (Astandard) was measured at room temperature 20 minutes later.

In the case of the registered method, a sample volume of about 20 mL was withdrawn from each vessel and filtered (pore size 0.45 µm). The absorption of the clear solutions at 239 nm (A239) was measured against simulated gastric fluid without pepsin [1] at room temperature, with a cell width of 10 mm. The amount of dissolved active substance in percent (%dissolved) was calculated using Formula 3, where 15,000 is a factor accounting for conversion to percent and dilution.

Formula 3:
%dissolved=A239*cstandard * 15,000/Astandard

For the fiber optic method the absorption was measured directly in the vessels at 239 (A239) and 450 (A450) nm with a path length of 10 mm. Formula 4 was used to determine the amount of active substance dissolved.

Formula 4:
%dissolved=(A239 -A450)*cstandard * 15,000/Astandard

Validation of the fiber optic method
The linearity was investigated in triplicate at five concentrations ranging from 25 to 125% dissolved (0.0017 ­ 0.0083 mg/mL). To spike the solutions, placebo powder was added before taking the absorption readings in the dissolution vessels under regular test conditions. The same data set was used to determine the accuracy as well as the precision. The fiber optic method was then compared with the validated and registered method. Six tablets from three lots were used. For reference, six vessels were filled with placebo-spiked model solutions corresponding to a dissolution of about 90% (0.0060 mg/mL). This experiment was carried out in replicates of six (n=6), filtering with a pore width of 0.2 µm. The resulting data sets enabled the accuracy as well as the precision to be calculated. The placebo response (six assays) and the dissolution profiles of six tablets were determined.

Results
Soft gelatin capsules
Method development

From Figure 2 it can be seen that the active substance does not show any absorption above 365 nm. The absorption of the placebo capsule is 0.050 over the entire wavelength range, owing to the turbidity. Above 365 nm the soft gelatin capsule absorption curve matches the turbidity line.

Figure 2: Absorption spectra of 0.26 mg/ml active substance (1), a soft gelatin capsule (2), and a placebo capsule (3) in 900 ml dissolution medium. .

For the new fiber optic method the shoulder of the active substance at 315 nm was chosen for quantification, just as in the registered model. The wavelength of 380 nm was chosen for the turbidity correction.

Validation of the fiber optic method
For the A (1%, 1 cm) calculation the six linearity assays were evaluated separately. The mean of the six A (1%, 1 cm) values was 52.96 cm-1 %-1 with a relative standard deviation of 0.51%.

To determine whether common acceptance criteria for linearity [3, 4] are met, the data sets were evaluated as a batch. The coefficient of correlation was 0.99968, exceeding the acceptance limit of > 0.99. The y-intercept met the acceptance limits as well, lying within the 95% confidence interval of 2% of the reference x-value (100% of active compound dissolved) around the origin. All group standard deviations as well as the relative repeatability standard deviation (0.44%) were within the acceptance criterion of <1.00%. The internal validation gave a mean recovery of 98.87%, satisfying the acceptance limits of 98.00 to 102.00%.

The method comparison data showed no significant difference (p = 95%). Thus the 95% confidence interval of the mean of the fiber optic method was entirely within the acceptance interval of 2.00% around the mean of the validated and registered method with filtration. The relative within-sample standard deviation with this data set was found to be 0.49%, satisfying the acceptance limit of <1.00%.

The placebo response was -0.17% for the placebo capsules and -0.35% for the soft gelatin capsule shells.

Robustness

Table 1: Influence of temperature on the detected percentage of dissolved active substance.

 Temperature [°C]

 Dissolution [%]

 25

 100.1

 30

 100.0

 35

 100.1

 40

 100.1

 45

 100.0
Table 1 demonstrates that the fiber optic detection method is unaffected by temperature (p = 95%).

Table 2: Dissolution values obtained at various probe immersion depths within the vessels. The depth of 3.0 cm was used for all other measurements.

Immersion depth [cm]

Assay

Vessel

0.5

2

3

4.5

1

1

102

102.2

102.1

102.1

2

102.1

102.1

102.1

102.1

3

101.6

101.9

101.5

101.9

4

102.1

102.2

102.1

102.2

5

101.8

101.9

102

101.7

6

102

101.9

102.1

102.3

2

1

99.7

99.8

99.7

99.5

2

99.4

99.3

99.5

99.4

3

99

98.9

98.8

98.9

4

101

101

101.1

100.8

5

99.6

99.5

99.5

99.2

6

99

99

98.9

99

Figure 3 shows the influence of paddle speed on the detected dissolution.
With the equipment used here the radial distance between the probe and the paddle could not be varied and was within the requirements of Ref. 1. Findings for probe window immersion depths of 0.5, 2.0, 3.0 and 4.5 cm are presented in Table 2.
A statistical comparison of the depths 0.5 and 4.5 cm did not show any significant differences (p = 95%).

Figure 3: Influence of paddle speed on the fiber optic quantification method. The dots correspond to the mean of twelve soft gelatin capsules and the vertical lines represent ± 1 standard deviation.

Furthermore, the absorption was measured (sixfold determination) in one vessel at six different immersion depths at 315 nm and 380 nm. The mean absorption readings at both wavelengths as well as their differences are presented with the corresponding standard deviations in Table 3 .

Table 3: The means of six absorption readings at 315 (A315) and 380 nm (A380) as well as the difference between the two (D A) are shown for different immersion depths in the same vessel. All obtained values are given with the corresponding standard deviations.

Immersion depth

A315

A380

A

[cm]

[AU]

[AU]

[AU]

1

1.2474 ± 0.0012

0.0757 ± 0.0003

1.1717 ± 0.0013

2

1.2474 ± 0.0012

0.0759 ± 0.0001

1.1715 ± 0.0012

3

1.2468 ± 0.0009

0.0757 ± 0.0002

1.1711 ± 0.0009

4

1.2472 ± 0.0017

0.0761 ± 0.0003

1.1711 ± 0.0017

5

1.2470 ± 0.0012

0.0755 ± 0.0003

1.1715 ± 0.0011

6

1.2471 ± 0.0005

0.0754 ± 0.0002

1.1716 ± 0.0005

These results show that the turbidity as well as the concentration of the active substance is homogeneous in the examined region of the vessel because the different immersion depth do not lead to significantly different results (p= 95%). Although the turbidity is measured during each quantification, further evidence about the robustness of the method is given owing to the small turbidity differences.

As can be seen in Figure 4, the mean dissolution profile has a sigmoid shape, with high variability between 6 and 15 minutes. The acceleration at around 6 minutes corresponds to the rupture of the capsule shell. After 15 minutes the entire liquid content of the soft gelatin capsule has escaped and the 100% plateau is reached.

Figure 4: Mean dissolution profile of six soft gelatin capsules. The vertical lines represent ±  1 standard deviation.

Benzodiazepine tablets
Method development

For the fiber optic method the maximum at 239 nm in Figure 5 was used for the determination of the active substance, just as in the registered method. Since the spectrum of the active substance meets the x-axis just before 450 nm, this wavelength was chosen for the turbidity correction. At the upper end of the spectrum the turbidity line is higher than the benzodiazepine tablet curve owing to the difference in particle size between the placebo powder turbidity and the turbidity resulting from tablet dissolution.

Validation of the fiber optic method
The linearity assays met the above-mentioned acceptance criteria concerning the coefficient of correlation (0.99993), the y-intercept, all the group standard deviations and the relative sample standard deviation (0.50%). The internal validation led to a mean recovery of 100.38%.

The results of the method comparison with tablets are shown in Table 4.

Table 4: Mean dissolution test results after quantification by the two methods. The standard deviation of the six readings is given in brackets.

Assay

Registered method [%]

Fiber optics [%]

1

91.5 (± 0.7)

92.3 (± 0.9)

2

88.7 (± 1.6)

88.5 (± 0.8)

3

87.9 (± 1.6)

87.9 (± 1.6).


The dissolution values differ by less than 2.00% and so meet the acceptance limits.
The method comparison with model solutions proved that the methods are equivalent: the 95% confidence interval of the mean of the fiber optic method lay fully within the acceptance interval of 2.00% around the mean of the registered method. The relative within-sample standard deviation determined with this data set was 0.60%, similarly meeting the acceptance criteria.

For the mean placebo response a value of -0.20% was found.
The benzodiazepine tablets show the typical dissolution profile (Figure 6) of an immediate release tablet formulation, characterized by a steep initial slope which gradually levels off, resulting in a final relative standard deviation of about 2%.

Figure 6. Mean dissolution profile of six benzodiazepine tablets. The vertical lines indicate the interval of ±1 standard deviation.

Discussion
With the benzodiazepine tablet model the spectra for the method development were not as smooth as in the case with the soft gelatin capsules since the particles causing turbidity were larger. While quantifying the amount of dissolved active substance, the influence of the turbidity and the turbidity particle size on the accuracy of the absorption reading was eliminated by accumulating 80 xenon lamp flashes in order to obtain one absorption reading. (The spectrometer used was equipped with a xenon flash lamp as light source.) This procedure was applied for the examination of both drug products.

The validation experiments for the two drug products were performed differently owing to the differences in quantification. In the case of the soft gelatin capsules the linearity assays involved more samples because they served to estimate A (1 %, 1 cm). The data range (5 to 120%) was chosen so as to optimally trace the sigmoid dissolution curve.
To compensate for the disintegration of the active substance provoked by UV light [5, 6], a one-point quantification with a reference solution was performed for each experiment. The standard solution for the one-point calibration is measured at room temperature, leading to a maximal theoretical deviation of 0.002 AU corresponding to the density decrease with increasing temperature [7, 8].

When examining the benzodiazepine tablets, the placebo powder turbidity differed from the tablet turbidity owing to the difference in particle size, as can be seen in the results of the method development experiments. For this reason the pore size of the filter had to be changed to 0.2 µm for the filtration of solutions containing placebo powder [9]. Because of the use of two different filter sizes, the method comparison was performed with tablets as well as model mixtures. For the model mixtures a dissolution of 90% was chosen, which is the regular value to be expected [9]. However, the method comparison with tablets required different acceptance criteria to deal with the scattering of the dissolution profiles, as can be deduced from the relative standard deviation of about 2% at the end of the mean benzodiazepine tablet dissolution curve [9]. In the case of the soft gelatin capsules, scattering was virtually undetectable at the end of the dissolution run owing to the design of this dosage form.

Although the robustness was examined with the soft gelatin capsules only, there is strong evidence that the robustness is the same in the case of the benzodiazepine tablets [10].

Conclusions
With the new method presented here the analysis time for manual dissolution testing can be drastically reduced. Thanks to the lack of sample preparation, the new fiber optic method is 6 to 8 times faster and avoids the use of disposable materials such as filters and syringes. The new method rapidly repays the necessary investments in the fiber optic probe and fiber optic coupler.

For automated dissolution testing, the filter and pump station can be replaced by fiber optic probes, thus eliminating the laborious qualification and validation of this equipment. Another advantage is the elimination of typical pumping and filtration problems such as air bubbles in the flow-through cells and pumping system, absorption by tubing, compatibility issues, volume deviations, filter clogging and drug hold-up.
Thus the use of fiber optics offers considerable advantages in manual dissolution testing and may be the ideal tool for automation of dissolution testing [11].

References
[1] USP 23 NF 18, United States Pharmacopeial Convention, Inc., Rockville, MD, (1995)
[2] Rohrs B.R., Stelzer D.J., Deaeration Techniques for Dissolution Media, Dissolution Technologies, 2 (2) 1-9 (1995)
[3] International Conference on Harmonization, Validation of Analytical Procedures: Methodology, ICH Harmonised Tripartite Guideline Q2B (1995)
[4] Pharma Switzerland, Quality Assurance and Quality Control, Validation of Analytical Methods, F. Hoffmann-La Roche Ltd., Basel (1998)
[5] Roth H.J., Eger K., Troschütz R., Arzneistoffanalyse, Gustav Fischer, Stuttgart, Jena, Lübeck, Ulm, 3 (1997)
[6] Debesis E., Revised Analytical Data Sheet for the Active Substance*, Red Corner Research Report No. N-33671, F. Hoffmann-La Roche Ltd., Nutley (1979)
Fiber Optic References...continued
[7] Stricker H. (ed.), Physikalische Pharmazie,
Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart, 3 (1987)
[8] Lide D.R. (ed.), Handbook of Chemistry and Physics, CRC Press, Boca Raton, New York, London, Tokyo, 76 (1996)
[9] Schatz C., Internal Communication about the Benzodiazepine Tablets, F. Hoffmann-La Roche Ltd., Basel (1999)
[10] Schatz C., Internal Communication about the Soft Gelatin Capsules, F. Hoffmann-La Roche Ltd., Basel (1999)
[11] Kostek L.J. et al., Automated Dissolution Testing Utilizing On-line Fiber Optic Probe UV Analysis, ISLAR '95 Proceedings, 355-367 (1995)

Correspondence:
Caspar Schatz
F. Hoffman-LaRoche, Ltd.
POBQ, Building 61/133
CH-4070 Basel Switzerland
Phone +41-61-687 14 33 Fax +41-61-688 8020
caspar.schatz@roche.com